31
Mount fixtures
Product Support:
Operation of HPS
The HPS is operated by controls on the HPS control box and the user control panel, and by functions
within the controlling software. Methods to start and stop the HPS are provided in
,
When the HPS is on, the hydraulic cylinder should be positioned so that the piston is neither fully
retracted nor fully extended. When the hydraulic cylinder is in either of these conditions, it is difficult for
the 59 Series controls to maintain the position and the servo loop control could build up enough error to
shut down the HPS.
Mount fixtures
Warning
The frame is supplied with multiple bolt patterns for the mounting of specimen load trains and testing
accessories. Some bolt patterns are designed for the mounting of specific standard Instron accessories
or fixtures, while others are more general for the mounting of a wide range of standard Instron
accessories or fixtures. Any bolt pattern can be used for the mounting of customer fixtures; appropriate
adaptation may be necessary. The capacity of any load train or accessory may be limited by the capacity
of the bolt pattern being used. Most bolt patterns can accommodate testing at full frame capacity. Refer
to the Note below.
Details of the bolt patterns provided on your frame can be found on the General Assembly drawing that is
located in the separately supplied “Reference Manual-Equipment”. General Assembly drawings are
model specific; please be sure to refer to the appropriate drawing for your model and its configuration.
Always wear appropriate personal protective equipment when preparing, operating and
maintaining this equipment. Personal protective equipment should include, but is not
limited to, eye protection and head protection. Other types of personal protective
equipment may also be needed. You must perform your own risk assessment and take
appropriate measures to protect yourself from harm.
Use of lifting or handling equipment (i.e. overhead crane, hoist, etc.) may be necessary to lift and install
testing accessories. Many testing accessories weigh in excess of 15 kgs (30 lbs) and may be difficult for
operators to install and remove from the frame. All of these considerations are the responsibility of the
customer.
Any bolt pattern that is six M10 x 1.5p tapped holes equally spaced on a 100 mm (3.937 in) bolt circle
diameter can only accommodate low range tension testing up to 100 kN (22,500 lbs).
Содержание 1000KPX-J3D
Страница 30: ...Chapter 3 Operation 30 M47 17035 EN Figure 5 User control panel...
Страница 50: ...Chapter 3 Operation 50 M47 17035 EN This page is intentionally blank...
Страница 68: ...Appendix CE Certificate 68 This page is intentionally blank...
Страница 69: ......
Страница 70: ...Product Support www instron com...