25
Heat load
Product Support:
Environmental conditions
lists the recommended environmental conditions in which the system should be operated and
stored.
Heat load
The HPS generates a certain amount of heat during operation. Refer to
for the approximate heat
output of the HPS operating at both a 10% duty cycle and 50% duty cycle. The duty cycle is defined as the
amount of time that the HPS operates under full load conditions. For example, operating at a 10% duty
cycle would mean performing approximately one 6-minute test at full load per hour.
Table 8. Recommended environmental conditions.
Operating Temperature:
+10 to +38 deg C (+50 to +100 deg F)
Storage Temperature:
-40 to +66 deg C (-40 to +150 deg F)
Humidity:
10% to 90% (non-condensing)
Atmosphere:
Designed for use under normal laboratory conditions. Protective measures may
be required if excessive dust, corrosive fumes, electromagnetic fields, or
hazardous conditions are encountered.
Table 9. Approximate heat output of various HPS models.
HPS Catalog Number
HPS Model
Approximate Heat Output at 10%
Duty Cycle
Approximate Heat Output at 50%
Duty Cycle
600KPX-D17
V22h
0.84 kW (2856 BTU/hour)
1.79 kW (6120 BTU/hour)
1000KPX-D18,
1500KPX-D18
V22e
0.12 kW (392 BTU/hour)
0.38 kW (1279 BTU/hour)
2000KPX-D19
V22f
0.12 kW (392 BTU/hour)
0.38 kW (1279 BTU/hour)
Содержание 1000KPX-J3D
Страница 30: ...Chapter 3 Operation 30 M47 17035 EN Figure 5 User control panel...
Страница 50: ...Chapter 3 Operation 50 M47 17035 EN This page is intentionally blank...
Страница 68: ...Appendix CE Certificate 68 This page is intentionally blank...
Страница 69: ......
Страница 70: ...Product Support www instron com...