Electromagnetic immunity - General
Directives and declaration of the manufacturer - electromagnetic emissions
The Apolus is intended to be used in the specific electromagnetic environment below. The user or customer of the
defibrillator should ensure that it is used in such an environment.
Immunity Test
Test Level
ABNT NBR IEC 60601
Compliance
Level
Electromagnetic Environment
Directives
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 KV contact
± 8 KV air
± 6 KV contact
± 8 KV air
Floors should be made of wood,
concrete or tiles. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC61000-4-4
± 2 KV for power supply
lines
± 1 KV for input/output
lines
± 2 KV for power supply
lines
± 1 KV for input/output
lines
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 KV differential mode
(phase - phase)
± 2 KV common mode
(phase - ground)
± 1 KV differential mode
(phase - phase)
± 2 KV common mode
(phase - ground)
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Drop:
0% UT for 0.5 cycles A
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.
0% UT for 1 cycle and
70% UT for 25/30 cycles
Monophasic A 0°.
Interruptions:
0% UT for 250/300
cycles.
Drop:
0% UT for 0.5 cycles A
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.
0% UT for 1 cycle and
70% UT for 25/30 cycles
Monophasic A 0°.
Interruptions:
0% UT for 250/300
cycles.
The quality of the power
supply should be that of a
typical commercial or hospital
environment. If the user of the
Apolus requires continued
operation during power interruption,
it is advisable that the Apolus is
supplied by an uninterrupted power
source or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at characteristic levels of
a typical commercial or hospital
environment.
NOTE U
T
is the AC mains voltage prior to application of the test level.