Field Calibration

The Micromate monitoring unit provides an option to calibrate the Sound Level 

Microphone in the field. A field calibration only uses one reference signal of 94 dB 

at 1 kHz. (The  Brüel & Kjaer Sound Calibrator Type 4231 is also capable of generating 

a 114 dB signal, make sure you set the device to 94 dB at 1 kHz.) A successful field 

calibration will store the values for Sound Pressure, Gain and Offset on the hardware 

of the microphone, if there is an error during calibration, no values are overwritten. 

In this case you can either try again or have the Sound Level Microphone calibrated 

at  Instantel  or  one  of  its  authorized  calibration  facilities.  Reports  generated  by 

the Micromate monitoring unit will specify the serial number, calibration date and 

calibration method. 
Example: 

Factory Calibration: 

UA12345, January 1, 2022 by Instantel

Field Calibration:  

UA12345, January 1, 2022 by Field Calibration

Note: Swapping Sound Level Microphones from one Micromate unit to another Micromate unit is permitted. As 

calibration values are stored on the Sound Level Microphone, this will not impact any Micromate unit settings. 

Field Calibration Procedure

  1.  Turn on the Micromate unit and connect the Sound Level Microphone, then press the 

Setup

 key on the unit.

  2.  Tap 

Maintenance Menu 

then scroll down and tap 

Calibrate Sound Level Microphone.

  3.  Follow the field calibration device’s operator instructions to properly join the microphone with the field  

        calibration unit. Set the reference to 94 dB at 1 kHz then turn the calibrator on. 
Note: When using either of the three 94 dB field calibrators, it is recommend that the ambient noise be at least 30 dB 

lower than this reference signal.  

For additional information please see 

https://www.pcb.com/Contentstore/mktgcontent/WhitePapers/

WPL_36_Acoustic_methods_calibration_PCB.pdf
  4.  Tap the 

Calibrate

 button on the Micromate monitoring unit screen.

  5.  Tap the 

Checkmark

 key to confirm the 

Proceed with calibration

 popup box. This will display

 Running Calibration

.

  6.  When the calibration is finished, the Micromate unit will display 

Field Calibration Complete

 along with the

        values for the measured 

Sound Pressure, Gain, and Offset

.

  7.  In the 

Field Calibration Complete 

confirmation box, tap the 

Checkmark

 key.

Important Notes:

• 

Only use an Instantel recommended calibration device (see the Tools and Materials Required section).

• Remove the microphone’s windscreen 

before

 attaching it to the field calibration device, as this can   

  impede proper calibration.

• Slide the Sound Level Microphone into and out of the field calibration device 

without

 twisting either end.

• Before reinstalling the Sound Level Microphone in the field, 

hand-tighten 

the microphone’s housing  

  extension and cap.

Calibration

The Sound Level Microphone 

must be sent

 to a manufacturer certified calibration facility, or Instantel headquarters 

for calibration. Please visit our calibration facility locator at www.instantel.com. Instantel recommends calibrating all 

Sound Level Microphones once a year and provides the Next Calibration dates on the unit and sensors. The factory 

calibration process ensures that the microphone is performing as it was designed and accurately measuring 

according to the IEC 61672 Class 1 standard. 
Using  a  manufacturer  certified  facility  ensures  the  proper  procedures  are  followed  when  calibrating  your 

microphone. This removes any uncertainty associated with the reliability and measurement accuracy of your 

microphone. It also helps ensure the integrity of the measured recorded events if they are ever called upon as 

evidence in a court of law. Using unauthorized facilities may have negative legal consequences if the reference 

equipment is not traceable, the proper procedures are not followed or if the certification documents are incomplete. 

Please visit our calibration facility locator at instantel.com.

Note: 

Field calibration is not 

as precise as factory 

calibration. 

The Field Calibration 

Procedure will change 

the date and 

overwrite

 

the factory calibration 

values.

The tracability will then 

be linked to the Field 

Calibration device.