Insize Wireless Digital Microscope User Manual
Insize
Co., Ltd. Page 28 of 32
:
Add text.
:
Undo.
:
Redo.
:
Calibration
: Ongoing operation.
:
Magnification for measurement
Calibration Manual
1.
Select a known measurement standard as the to-be-observed object, such as a ruler, to be used for calibration.
Bring image of the selected ruler into focus by turning the “Focus & Magnification adjustment wheel”,
determine the magnification by taking a reading from the scale on the focus dial, enter the scale value in the
“Scale” box in the upper right corner of the screen, and then take the image.
2.
Now double click to open the image of the measurement calibration standard, and the scale of magnification
will be automatically loaded for measuring.