insize ISHR-D121 Скачать руководство пользователя страница 3

3

4

Main Technical Parameters 

Initial test force: 10kg
Test force: 30, 100, 150kg
Maximum magnetic force: 350kg
Force applying method: by screw 
Indenter: 120°diamond cone, Ø1.588mm carbide balls 
Testing Items: HRC, HRB, HRA 
Resolution: 0.1HR 
Accuracy: Meets the requirements of ISO6508/6506 
Application:  All  metals,  including  steel,  forged  steel,  cast  iron,  copper, 
copper alloy, aluminum, aluminum alloy and carbides etc.
Flat workpiece requirement: length>190mm, width>60mm, thickness>5mm
C y l i n d e r   o r   t u b e   w o r k p i e c e   r e q u i r e m e n t :   l e n g t h > 2 0 0 m m ,  
diameter>Ø50mm, wall thickness>8mm  

Panel And Screen

Operation

Test Conditions:
Make sure the surface is flat and smooth. It is forbidden to have scale, 
decarburized layer, steel pit or dust on the surface of the part.

1

”:  Turn

 

on

 

and

 

off

 

the

 

power.  After  3

 

minuets

 

the

 

instruments

 

without using, the

 

power

 

will

 

be

 

off

 

automatically.

ON

”: After  pressing  it,  the  instrument  is  on  status  setting  and  the 

menu  displaying,  then  press  "Manu"  again,  it  would  be  on  next  level 
menu.

MENU

”: It’s a multi-functional button. On status of test, press it realizing 

the  function  of  counting;  on  menu  setting,  it  could  move  the  arrow 
upward; on the status of calibration, it could increase values.

STAT

”: It’s a multi-functional button. On status of test, press it to see 

the  conversion  value;  on  menu  setting,  it  could  move  the  arrow  down 
ward; on the status of calibration, it could decrease values.

CONV

”: It’s a double functional button. On status of test, it could delete 

current  values;  on  status  menu  setting,  it  could  confirm  the  selection 
and back to previous menu.

DEL

Note:
Make sure the indenter is higher than the bottom surface before put the 
instrument  on  workpiece,  the  indenter  can’t  touch  the  test  part. 
Otherwise, the indenter maybe crash to damaged.

The  tester  should  be  attached  to  test  part  flatly  and  steadily  with  the 
indenter erected to the surface of test part. Turn the magnetic switch to 
"On", keeping the tester fixed onto the test part.

2

Power  on  the  instrument,  set  the  dwell  time  and  select  the  test  scale 
HRA, HRB or HRC. The scale will display on the screen.

3

Rotate  the hand  wheel  to load  test force steadily, observe  the screen, 
stop or add test force extremely slowly when test force is close to the  
specified force. Try to keep the test force, now the time counting down 
displays  on  screen.  When  the  time  counting  down  finished,  spin  hand 
wheel  reversely  to  release  test  force  steadily  until  hardness  values 
display. Then one full operation is accomplished.

4

Test force

60kg

100kg

150kg

Scale

HRA

HRB

HRC

scale

indication calibration

battery indicator

test force

dwell time

unit

button

Отзывы: