High Priority Alerts

The following high priority alerts  are accompanied by a five beep pattern, repeated every  

10 seconds and a flashing yellow light.

Description

Display  Icons

Condition/Action/Explanation

Battery Empty 
Attach Plug

Concentrator has insufficient battery power to 

produce oxygen. Attach external power supply 

or exchange battery, then restart unit if  

necessary by pressing On/Off button.

Battery HOT

Battery has exceeded temperature limit while 

concentrator is running on battery power.  

Concentrator has stopped producing oxygen.  

If possible, move concentrator to a cooler  

location, then turn power off and back on.  

Ensure air intake and outlet vents have clear 

access and particle filters are clean. If condition 

persists, switch to a backup source of oxygen and 

contact your equipment provider.

System HOT

Concentrator temperature is too high and 

oxygen production is shutting down. Ensure air 

intake and outlet vents have clear access and 

particle filters are clean. If condition  

persists, switch to a backup source of oxygen 

and contact your equipment provider.

System COLD 

This may result from the concentrator being 

stored in a cold environment (below 0ºC (32ºF)). 

Move to a warmer environment to allow the 

unit to warm up before starting it. If condition 

persists, switch to a backup source of oxygen 

and contact your equipment provider.

System Error

Concentrator has stopped producing oxygen 

and is shutting down. You should: 

1. Switch to backup oxygen source 

2. Contact your equipment provider

6. Troubleshooting

Problem

Possible Cause

Recommended Solution

Any problem accompanied 

by information on  

concentrator display,  

indicator lights and/or 

audible signals

Refer to Chapter 5

Refer to Chapter 5

Содержание One G5

Страница 1: ...1 USER MANUAL ...

Страница 2: ...2 ...

Страница 3: ...es 5 Input Output Connection 6 Chapter 4 Operating Instructions 6 General Instructions 6 Power Supply Options 8 Inogen One G5 Accessories 10 Traveling with Inogen One G5 12 Chapter 5 Audible and Visible Signals 12 Chapter 6 Troubleshooting 17 Chapter 7 Cleaning Care and Maintenance 18 Cannula Replacement 18 Inogen One G5 Column Change Procedure 19 Chapter 8 Specifications 22 Disposal of Equipment ...

Страница 4: ...as determined this device conforms to all applicable FAA requirements for POC carriage and use on board aircraft Compliant with the Waste Electrical and Electronic Equipment Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment WEEE RoHS recycling directive Keep Dry Indoor or Dry Location Use Only Do Not Get Wet AC Power DC Power Refer to instruction manual ...

Страница 5: ...n traveling Inogen assumes no liability for persons choosing not to adhere to manufacturer recommendations Cautions and Warnings Cautions A caution indicates that a precaution or service procedure must be followed Disregarding a caution could lead to a minor injury or damage to equipment Additional monitoring or attention may be required for patients using this device who are unable to hear or see...

Страница 6: ...t the tubing is not kinked or pinched in any way Replace the nasal cannula on a regular basis Check with your equipment provider or physician to determine how often the cannula should be replaced The Inogen One G5 is designed to provide a flow of high purity oxygen An advisory alert Oxygen Low will inform you if oxygen concentration drops If alert persists contact your equipment provider Ensure th...

Страница 7: ...r physician immediately Ensure that the automobile power socket is adequately fused for the Inogen One G5 power requirements minimum 15 Amp If the power socket cannot support a 15 Amp load the fuse may blow or the socket may be damaged The tip of the Cigarette Adapter Plug becomes HOT when in use Do not touch the tip immediately after removal from an auto cigarette lighter socket It is the respons...

Страница 8: ... bell icon flashing yellow light and display message when the alert is enabled If power is lost the breath detection audible alert remains set in the user preferred mode User Interfaces Item Description Function 5 Display This screen displays information regarding flow setting power status battery life and errors Display appearance will vary Before use remove the static cling FCC label from the sc...

Страница 9: ... and exhaust must have clear access Locate the Inogen One G5 in such a way that any auditory alerts may be heard Always operate the Inogen One G5 in an upright position see image for proper orientation 3 Ensure particle filters are in place at both ends of device 4 Insert the Inogen One G5 battery by sliding battery into place until the latch returns to the upper position 5 Connect the AC input pl...

Страница 10: ...ime may be needed if your Inogen One G5 has been stored in extremely cold temperatures 8 Set the Inogen One G5 Concentrator to the flow rate prescribed by your physician or clinician Use the or setting buttons to adjust the Inogen One G5 to the desired setting The current setting can be viewed on the display 9 Position the nasal cannula on your face and breathe through your nose The Inogen One G5 ...

Страница 11: ...attery will discharge The display will indicate the estimated remaining percentage or minutes of use When the concentrator detects that the battery life is low with less than 10 minutes remaining a low priority alert will sound When the battery is empty the alert will change to a high priority When battery life is low do one of the following Plug the Inogen One G5 into an AC or DC power source usi...

Страница 12: ... 75 to 100 full 3 LEDs Light 50 to 75 full 2 LEDs Light 25 to 50 full 1 LED Lights 10 to 25 full 1 LED Blinks Battery is less than 10 full and needs to be recharged Power Supply Overview The Inogen One G5 AC power supply BA 501 is used to power the Inogen One G5 concentrator from an AC power source The Inogen One G5 AC power supply is specifically designed for use with the Inogen One G5 Oxygen Con...

Страница 13: ...le or double battery and zip up the bottom flap 2 The cannula barb will be exposed at the top of the bag for proper attachment The display screen can be accessed by grabbing the short material pull tab at the top flap just above the grab handle on the back top section of the bag 3 Both intake vents should be visible through the open mesh panels on both sides of the bag The exhaust vent should be v...

Страница 14: ...e call Inogen Client Services External Battery Charger BA 503 The Inogen One G5 external battery charger will charge the Inogen One G5 single and double batteries 1 Plug the External Battery Charger AC power supply cord into an electrical outlet 2 Plug the External Battery Charger AC power supply into the battery charger 3 Slide your charger onto the Inogen One G5 Battery by clicking and locking i...

Страница 15: ... class of service You must check with your airlines for availability and any specific requirements for battery life duration 48 hours before traveling In this case the following procedure regarding transition from battery power to aircraft electrical power must be followed º Remove the battery from the Inogen One G5 º Connect the DC power plug to the Inogen One G5 and plug into available airline p...

Страница 16: ...arge level less than 10 The Inogen One G5 is operating from an external power source with no battery present Mode Icons These are the mode icons shown in the display s window The breath detection audible alert has been enabled The breath detection audible alert is disabled This is the default condition Sound Level 1 Sound Level 2 Sound Level 3 Sound Level 4 Display Icons The icons below are exampl...

Страница 17: ... used for oxygen production It is normal to see a fully charged battery read between 95 and 100 when external power is removed This feature maximizes the useful life of the battery Sieve Reset Displayed when column maintenance is required and once the replacement columns have been installed Sieve Reset Success Displayed once the columns have been successfully reset Notification Icons The Inogen On...

Страница 18: ...e Check Battery Battery error has occurred Check the connection of your battery and ensure that it is properly attached and latched on concentrator If battery error recurs with same battery stop using the battery and switch to a new battery or remove battery and operate concentrator using external power supply Oxygen Low Concentrator is producing oxygen at a slightly low level 82 for a period of 1...

Страница 19: ...ed to concentrator there are no kinks in tubing and cannula is positioned properly in your nose Oxygen Error Oxygen output concentration has been below 50 for 10 minutes If condition persists switch to your backup oxygen source and contact your equipment provider to arrange for service O2 Delivery Error A breath has been recognized but proper oxygen delivery has not been detected Battery HOT Warni...

Страница 20: ...up source of oxygen and contact your equipment provider System HOT Concentrator temperature is too high and oxygen production is shutting down Ensure air intake and outlet vents have clear access and particle filters are clean If condition persists switch to a backup source of oxygen and contact your equipment provider System COLD This may result from the concentrator being stored in a cold enviro...

Страница 21: ...r physician and or equipment provider and or cannula manufacturer s instructions for replacement information A single lumen cannula up to 25 feet in length is recommended to ensure proper breath detection and oxygen delivery Case Cleaning You may clean the outside case using a cloth dampened with a mild liquid detergent such as DawnTM and water Filter Cleaning and Replacement The particle filters ...

Страница 22: ... battery BA 500 Inogen One G5 double battery BA 516 Carry Bag CA 500 Backpack CA 550 External Battery Charger BA 503 AC Power Supply BA 501 DC Power Cable BA 306 Maintenance Items Replacement intake particle filters RP 500 Output Filter Replacement Kit RP 404 Inogen One G5 columns RP 502 Note Additional options may be available for country specific power cords To order contact Inogen or your equip...

Страница 23: ...pletely from the Inogen One G5 Both columns are removed as one piece 8 Column metal tube Installation Remove dust caps of new column assembly Make sure there is no dust or debris where the dust caps were located 9 Insert column assembly into the Inogen One G5 concentrator Do not leave the column ends exposed column assembly should be inserted into the Inogen One G5 as soon as the dust caps have be...

Страница 24: ...ttons on the device s screen or within Inogen Connect App Steps from your device s display º º Press and hold the plus and minus button for 5 seconds The screen will display the following informational icon Release button once icon is displayed on screen º º Press the alert button once and screen will display the following informational icons º º Press the power button to turn on the Inogen One G5...

Страница 25: ...ry Battery Charging Time Up to 4 hours for a single battery Up to 8 hours for a double battery Environmental Ranges Intended for Use Temperature 41 to 104 F 5 to 40 C Humidity 0 to 95 non condensing Altitude 0 to 10 000 ft 0 to 3048 meters Environmental Ranges Intended for Shipping and Storage Temperature 13 to 158 F 25 to 70 C Humidity 0 to 95 non condensing Store in a dry environment Altitude 0 ...

Страница 26: ...of in unsorted municipal waste Within Europe contact the EU Authorized Representative for disposal instructions The battery contains lithium ion cells and should be recycled The battery must not be incinerated Inogen One G5 Pulse Volumes at Flow Settings Inogen One G5 Flow Setting Breaths per Minute 1 2 3 4 5 6 15 14 0 28 0 42 0 56 0 70 0 84 0 20 10 5 21 0 31 5 42 0 52 5 63 0 25 8 4 16 8 25 2 33 6...

Страница 27: ...s RTCA DO 160 Note IT network is a system composed of wireless Bluetooth transmission between the Inogen One G5 and the Inogen Connect Application Connection of the Inogen One G5 to an IT Network could result in previously unidentified risks to patients operators or third parties Subsequent changes to the IT network could introduce new risks and require additional analysis Changes to the IT networ...

Страница 28: ...occur in the vicinity of equipment marked with the following symbol Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst EC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV f...

Страница 29: ...mum output power of the communications equipment 0 01 0 1 1 10 100 Separation Distance According to Frequency ofTransmitter M Rated Maximum Power Output of Transmitter W For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output p...

Страница 30: ...27 Inogen Inc 326 Bollay Drive Goleta CA 93117 Toll Free 877 466 4362 1 805 562 0515 Outside the USA E mail info inogen net www inogen com 2019 Inogen All rights Reserved PN 96 08649 00 01 B ...

Отзывы: