Low Priority Alerts
The following low priority alert messages are accompanied by a
double beep
and a
solid red light
.
Condition/Action/Explanation
External power supply is too low to run the unit, which
is running on battery power despite being plugged in.
Check external power supply connections. If condition
persists, contact your equipment provider.
External power supply is too high. Check external power
supply connections. If condition persists, contact your
equipment provider.
Battery power is low, with less than 10% remaining. Attach
external power supply or power down and insert a fully
charged battery.
Battery error has occurred. Switch to a new battery or
remove battery and operate concentrator using external
power supply. If battery error recurs with same battery,
stop using the battery and contact your equipment
provider.
Concentrator is producing oxygen at a slightly low level
(<82%). If condition persists, contact your equipment
provider.
Message Display & Text
External
Power Low
External
Power High
Battery Low
Attach Plug
Battery Error
See Manual
Oxygen Low
See Manual
24
Low Priority Alerts (continued)
Condition/Action/Explanation
Battery has exceeded its charging temperature and charging
has stopped. The battery will not charge while this alert
is present but will begin to charge when the battery
temperature returns to the normal operating range.
If battery charging is desired sooner, remove the battery
from the concentrator and allow it to cool in an open area for
approximately 10-15 minutes. Then, re-insert the battery into
the Inogen One® G2. If the problem still persists, contact your
equipment provider.
Concentrator is producing oxygen but cannot report
battery status. Replace battery. If condition persists,
contact your equipment provider.
The concentrator requires servicing at the earliest
convenience. The concentrator is operating to specification
and may continue to be used. Contact your equipment
provider to arrange for service.
The concentrator’s oxygen sensor has malfunctioned.
You may continue to use the concentrator. If the condition
persists, contact your equipment provider.
Message Display & Text
Remove Battery
to Cool
Comm Error
See Manual
Service Needed
02 Sensor Fail
See Manual
Audible and Visible Signals Chapter 4
25
English