![Inogen IO-200 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/inogen/io-200/io-200_technical-manual_2064280010.webp)
CONFIDENTIAL and
PROPRIETARY
96-02735-00-01 revC
INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 10 of 20
o
Replacement product filters (RP-101)
o
USB storage drive for data log transfers (such as PNY PN# P-FD2GBATT2-SF)
NOTE
During a normal field maintenance visit, the technician may elect to turn off the
concentrator for approximately 30 minutes. If the patient requires oxygen during
this period, Inogen recommends making arrangements to bring an extra oxygen
supply (such as a supplemental Inogen One G2 Concentrator).
3.1.4.
System Inspection
At the start of any maintenance visit:
1.
Be sure to ask the patient if they have experienced any difficulties in operating
the equipment.
2.
Be sure to ask the patient if they have observed any malfunctions or changes in
characteristics of the equipment.
3.
Visually inspect the device, batteries, and accessories for cracks or other
damage.
4.
Feel the sides of the device for vibration and listen for unusual noises, rattles, or
other signs that the device requires service.
CAUTION
Discovery of cracks or other types of external damage may be indicative of other
internal damage that may not be visible. If such external damage is discovered, be
certain to inquire as to how it occurred, and whether any changes in the device
have been noticeable since its occurrence. If you have any concern over the safety
of the device, arrange for equipment servicing.
3.1.5.
Product (Bacterial) Filter
Oxygen concentrator safety standards (ISO 8359) require use of a filter in the product
line capable of removing particles larger than 10 microns. This filter is intended to
protect the user from bacterial infection, and from inhalation of small particles in the
product gas flow. The Inogen One G2 includes a product filter, conveniently located
behind the removable cannula nozzle fitting. Inogen suggests that this filter be
replaced between patients.
AMSR
oxygenconcentratorstore.com
1-877-774-9271