DCR 20-328
INOGEN AT HOME OXYGEN
CONCENTRATOR TECHNICAL MANUAL
Page 9 of 15
5. Press the column latch button with thumb or finger and pull the column out
of the concentrator using the pull ring attached to the column cap.
6. Remove the column completely from the concentrator.
7. Repeat steps 5-6 to remove the other column.
Installation of the new columns:
8. Remove the upper and lower dust caps from each column. There should be
a total of two dust caps removed from each column. Make sure there is no
dust or debris where the dust caps were located.
9. Insert the new column into the concentrator.
10. The column should click into place and the spring loaded latch button
should fully return to the closed position when the column is completely
inserted.
11. Repeat steps 9-10 to install the other column.
12. Replace the lower cover and return the concentrator to its upright position.
13. Connect the power supply cord from the concentrator and plug into an
electrical outlet.
Do not power on
the concentrator.
14. Press and hold the plus (+) button for ten seconds. Flow indicators 1, 3 & 5
will be illuminated and columns will be reset. Release the plus button.
15. Press the power button to turn on the concentrator and use normally.
3.2. Maintenance by the Patient
3.2.1. Cannula Replacement
The nasal cannula should be replaced on a regular basis. A single lumen cannula of 4
to 100 feet in length must be used. Inogen has certified its performance data with the
Salter Laboratories 16SOFT cannula.
CAUTION
Do not use total cannula tubing length exceeding 100 feet with the Inogen
At Home unless proper saturation has been verified by a clinician.
3.2.2. Particle Filter Cleaning and Replacement
On the side of the Inogen At Home, ambient air passes through a particle filter that
removes dust fragments. This particle filter must be cleaned on a weekly basis to
CAUTION
Do not leave the column exposed after the dust caps are removed; it
should be inserted into the concentrator as soon as the dust caps have
been removed to minimize environmental exposure and maximize
lifespan.
NOTE
Column change instructions are only to be used when maintenance is
required. Column should only be removed during this maintenance
procedure.