Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d=1.2√
P
150 kHz to 80 MHz
d=1.2√
P
80 MHz to 800 MHz
d=2.3√
P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
3 Vrms
150 kHz
to 80 MHz
3V/m
80 MHz
to 2.5 GHz
3 Vrms
3V/m
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Compliance
Level
IEC 60601
Test Level
Immunity Test
Electromagnetic
Environment - Guidance
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
The Concentrator is intended for use in the electromagnetic environment specified below. The user
of the Concentrator should make sure it is used in such an environment.
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
EC 61000-4-4
Surge
IEC 61000-4-5
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s)
to earth
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the [ME EQUIPMENT or
ME SYSTEM] requires continued operation during power mains
interruptions, it is recommended that the
[ME EQUIPMENT or ME SYSTEM] be powered from
an uninterrupted power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital or home
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
<5%
U
T
(>95%
dip in
U
T
) for
0.5 cycle
40%
U
T
(60%
dip in
U
T
) for
5 cycles
70%
U
T
(30%
dip in
U
T
) for 25
cycles
<5%
U
T
(>95%
dip in
U
T
) for
5 sec
<5%
U
T
(>95%
dip in
U
T
) for
0.5 cycle
40%
U
T
(60%
dip in
U
T
) for
5 cycles
70%
U
T
(30%
dip in
U
T
) for 25
cycles
<5%
U
T
(>95%
dip in
U
T
) for
5 sec
3 A/m
3 A/m
Содержание at home
Страница 1: ...1 user manual ...
Страница 2: ...2 ...
Страница 19: ...16 ...
Страница 20: ...17 ...
Страница 21: ...18 benutzerhandbuch ...
Страница 22: ...19 ...
Страница 39: ...36 ...
Страница 40: ...37 ...
Страница 41: ...38 gebruikershandleiding ...
Страница 42: ...39 ...
Страница 59: ...56 ...
Страница 60: ...57 ...
Страница 61: ...58 manuel d utilisation ...
Страница 62: ...59 ...
Страница 79: ...76 ...
Страница 80: ...77 ...
Страница 81: ...78 manual del usuario ...
Страница 82: ...79 ...
Страница 99: ...96 ...
Страница 100: ...97 ...
Страница 101: ...98 manuale utente ...
Страница 102: ...99 ...
Страница 119: ...116 ...
Страница 120: ...117 ...
Страница 121: ...118 manual do utilizador ...
Страница 122: ...119 ...
Страница 139: ...136 ...
Страница 140: ...137 ...
Страница 141: ...138 ...