Troubleshooting

EMC Information

Diagnosis

Guidance and manufacturer's declaration - electromagnetic emissions

Guidance and manufacturer's declaration - electromagnetic immunity

The screen shows 'Erl'

Buttons have no

response

Emissions test

Immunity test

Compliance level

Electromagnetic environment -guidance

IEC60601 

test level

Compliance

Electromagnetic environment – guidance

No display or 

abnormal display

Power off when open

The screen shows

 'Lo' or 'Hi'

Solution

1. Check the object you are attempting to measure. The device is unable to effectively penetrate hair, 

water and cosmetics.

2. Check the environment as any fluctuating temperatures will influence the reading.

3. Have you just used the device to read very high or low temperature? Then leave the device for 10 minutes.

4. The temperature you are attempting to read is greater than 42.9°C or lower than 32°C.

This device is intended for using in the electromagnetic environment specified below. You should ensure it is 

used in such an environment.

RF emissions CISPR 11

This device is intended for using in the electromagnetic environment specified below. You 

should ensure it is used in such an environment.

Electrostatic discharge 

(ESD) IEC 61000-4-2

±2,±4,±6kV for 

contact 

discharge 

±2,±4,±8kV air 

discharge

±2,±4,±6kV for contact 

discharge ±2,±4,±8kV 

air discharge

Floors should be wood, concrete or ceramic 

tile. If floors are covered with synthetic 

material, the relative humidity should be at 

least 30%.

Mains power quality should be that of a 

typical commercial or hospital environment.

Mains power quality should be that of a 

typical commercial or hospital environment.

Mains power quality should be that of a 

typical commercial or hospital environment. 

If the user of this device requires continued 

power without interruption we recommend 

this device is powered from an 

uninterruptible power supply or a battery.

Power frequency magnetic fields should be 

at levels characteristic of a typical location 

in a typical commercial or hospital 

environment.

N/A

N/A

N/A

3 A/m

±2kV for AC 

power lines 

±1kV for DC 

power lines

±1kV line(s) to 

line(s)

±2kV line(s) to 

earth

<5% Ur (>95% 

dip in Ur) for 0.5 

cycle                      

<40% Ur (>60% 

dip in Ur) for 5 

cycles                    

<70% Ur (30% 

dip in Ur) for 25 

cycles                   

<5% Ur (>95% 

dip in Ur) for 5s

3 A/m

Electrical fast 

transient/burst IEC 

61000-4-4

Voltage dips, short 

interruptions and voltage 

variations in power supply 

input lines IEC 61000-4-11

Power frequency (50/60Hz) 

magnetic field IEC 

61000-4-8

Note: Ur is the AC mains voltage prior to application of the test level.

Surge IEC 61000-4-5

RF emissions CISPR 11

Class B

N/A

N/A

Harmonic emissions IEC 61000-3-2

Voltage fluctuations/flicker 

emissions IEC 61000-3-3

Group 1

This device uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.

This device is suitable for use in all establishments other than 

those directly connected to the public low-voltage power supply 

network that supplies buildings used for domestic purposes.

The environment temperature exceeds the normal working range of 16°C – 35°C (60.8°F - 95°F)

1. Load and unload the battery
2. Check if the setting is in progress

Unload the batteries and load again

Check the batteries, load and unload the batteries again

Содержание INH-GB-01

Страница 1: ... HEALTH NON CONTACT INFRARED THERMOMETER INH GB 01 USER MANUAL innovuhealth co ...

Страница 2: ...ccuracy with age so we advise replacing it Product Intended Use The infrared thermometer is intended for the measurement and monitoring of human body temperate for people of all ages Attentions Notes Before use please read the instructions and make sure the batteries provided have been installed Ensure this device does not have contact with liquid or extreme high low temperature Avoid this device ...

Страница 3: ...unction default alarm value is 38 0 C and above Capacity to store 20 readings in memory Optional unit of temperature C or F Automatic shut down and power saving Measurement Method Icon Definition Measuring mode 1 Measuring mode 2 Temperature unit Object temperature mode Battery capacity Icon Icon Definition Forehead and object temperature measurement Ear temperature measurement Celsius or Fahrenhe...

Страница 4: ...e MEM button and hold for 6 seconds and the screen will then display C If you want to change it to F press the MEM button once more and hold for 3 seconds C Voice switch setting Whenpoweredon holdtheMEMbuttonfor4seconds itwillbeeponceandthencloseoropenup Note thesoundkeycanbeseton offdependingontheenvironment e g itcanbeoffwhensilenceisappreciated D Shutdownkey Whenpoweredon holdtheMEMbuttonfor8se...

Страница 5: ...The thermometer can then begin taking object measurements by pointing the device at the object at 3 5mm away Press and hold the MEM button When you hear a beep the temperature will then display on the screen The device only needs routine cleaning and period or times can be decided according to user s need The surface cleaning use a clean soft cloth or cotton swab stick moistened with a little medi...

Страница 6: ...ains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of this device requires continued power without interruption we recommend this device is powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typi...

Страница 7: ...nt should be used no closer to any part of the device than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P NOTE 1 At 90MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection fr...

Отзывы: