Innovo INV-DC200 Скачать руководство пользователя страница 7

11 

 

1.Applicable 
regulations and 
laws 

ASTM 1965/EN12470-5/GB/T 19146-2010 

2.Temperature units  °F/°C, adjustable                                                                                                                                                             

3.Measurement 
range 

Forehead temperature mode: 89.6

°

F – 107.9 

°

F/32.0°C-42.2°C 

ear temperature mode:/32.0

°

F – 212.0 

°

F/ 

0°C-100.0°C     

4.Precision   

±0.4

°

F/±0.2°C 

5.Display resolution  0.1°F/0.1°C 

6.Latency Time 

1 second 

7.Abnormal state 
display 

LCD displays “L °C” if the measured temperature is 
below the minimum of measurement range. 
LCD displays “H °C” if the measured temperature is 
below the minimum of measurement range. 
LCD displays “Er1”if the temperature measurement 
circuit is abnormal (fault of SENSOR or the 
temperature measurement circuit), or the ambient 
temperature exceed 50°F-104°F (10°C~40°C). 
There shall be “ErC” if the calibration process is not 
completed or EEPROM is abnormal. 

8.Sound 

volume

50 db (the perpendicular distance from dB 

Volume sensor to thermometer is 10cm )   

9.Automatic 
shutdown function 

10s±1s 

10.Low-voltage 
display function 

The product shall display low-voltage signal if the 
voltage is below 2.51V±0.15V. 

11.Memory function  Memorize 20 groups of measured temperature. 

12.Current 
consumption 

I

stand-by

<2µA; I

working

<0.5mA; I

Buzzer on

<2mA

I

banklight

<15mA 

13.LED backlit 
specifications 

1.2cd/m

2

                                                       

14.Operational 
conditions 

ASTM (50°F-104°F)10°C-40°C /15-95%RH   

15.Type of 
measuring 

Applicable for forehead temperature and ear 
temperature 

 

The normal body temperature 
range slightly varies with age 
and gender. Generally, 
newborns or children have 
higher body temperatures than 
adults, and adults have higher 
body temperatures than the 
elderly. Women's body 
temperatures are 
approximately 0.3°C (0.5 °F) 
higher than men's. 

3. 

Body Temperature 

 
You can take a body temperature via the forehead, ear canal, armpit, 
mouth or rectum. The temperature measured at different parts of the 
body may differ slightly. 

Body Part    Normal Temperature Range 

 

Forehead 

96.44°F-100.04°F 
(35.8°C–37.8°C) 

Ear canal 

96.44°F-100.4°F 
(35.8°C–38.0°C) 

Mouth 

95.9°F-99.5°F   
(35.5°C–37.5°C) 

Armpit 

94.46°F-99.14°F 
(34.7°C–37.3°C) 

Rectum 

97.88°F-100.4°F 
(36.6°C–38.0°C) 

 

Variation in body temperature 

 

Normal body temperature 
varies by the time of day and is 
also affected by external 
factors. 
The body temperature of an 
individual is the highest 
between 2:00 a.m. and 4:00 
a.m. and the lowest between 
2:00 p.m. and 8:00 p.m. An 
individual's body temperature 
typically changes by less than 
1°C (1.8 °F) each day. 

Содержание INV-DC200

Страница 1: ...INV DC200 Dual Mode Digital Infrared Thermometer Instruction Manual Ver 2 1 0482 ...

Страница 2: ...eter simply scan the QR code above or visit http innovogroups com products innovo thermometer Introduction Thank you for purchasing the Innovo INV DC200 Dual Mode Digital Infrared Thermometer INV DC200 has undergone rigorous clinical tests and has proven to be safe reliable and accurate when used in accordance with the operation manual We wanted to make a thermometer that will accompany your child...

Страница 3: ...DC200 1 2 Precautions Care and Maintenance 2 3 Product Design 5 4 Measuring forehead temperature 5 5 Measuring ear temperature 5 6 Instructions to read and modify display 5 7 Troubleshooting 7 8 Technical Specifications 9 9 After sale service 10 10 Symbols 10 ...

Страница 4: ...ogy precise and immediate measurements Safe and hygienic Unlike traditional thermometers there is no glass or mercury that could pose as a potential health hazard The thermometer is made up of ABS and TPR plastics a temperature sensor an Infrared temperature measuring element a microcomputer controlled circuit and a LCD screen BPA and latex free Memory Recall Has a Memory Mode that can recall 20 r...

Страница 5: ...ot use the thermometer for purposes not specified in this User Manual Follow the instructions stated herein and operate the thermometer carefully when measuring children s temperature The device is not designed to be used for newborn babies The device is not a continuous monitoring device Do not use the thermometer under an ambient temperature higher than 40 C 104 F or lower than 10 C 50 F which i...

Страница 6: ... plastic sheet Batteries should be installed according to the stated polarities Otherwise the device bracket might be damaged Please remove the batteries if you are not planning to use the thermometer for a long time Warnings This is not a toy Please keep the Dual mode infrared thermometer out of children s reach The infrared Dual mode infrared thermometer is not a replacement for seeking medical ...

Страница 7: ...ature 12 Current consumption Istand by 2µA Iworking 0 5mA IBuzzer on 2mA Ibanklight 15mA 13 LED backlit specifications 1 2cd m2 14 Operational conditions ASTM 50 F 104 F 10 C 40 C 15 95 RH 15 Type of measuring Applicable for forehead temperature and ear temperature 4 The normal body temperature range slightly varies with age and gender Generally newborns or children have higher body temperatures t...

Страница 8: ... cover and insert the sensor into the ear canal 10 Symptom Possible Cause Solution symbol is displayed after the thermometer powers on low same model or specifications Er1 is displayed The ambient temperature is lower than 10 C 50 0 F or higher than 40 C 104 F Take a measurement under an ambient temperature between 10 C 50 0 F and 40 C 104 F The temperature reading is lower than the typical body t...

Страница 9: ...are correctly installed according to the polarity symbols in the battery compartment The thermometer is faulty If the warranty period does not expire contact Innovo Only the battery The battery level is Use new batteries of the 6 Incorrect position Correct Position Figure 1 For children 1 year old gently pull the ear up to ensure the ear canal is straight See Figure 1 Position the temperature prob...

Страница 10: ...seconds This indicates that the subject may have a fever Please consult your doctor if you are not sure or or Forehead Measurement mode Eardrum measurement mode The measured value exceeds 107 9 F 42 2 C 7 short beeps and the display will flicker with a red backlit for 3 seconds 8 The measured value falls below 89 6 F 32 C 7 short beeps and the display will flicker with a red backlit for 3 seconds ...

Отзывы: