U Quick Start Guide
Copyright © Innovative Technology Ltd 2018
Doc: ICU Quick Start Guide
Version: 1.2
Page 9 of 17
3
MACHINE INTEGRATION
3.1
Power Specification
The unit shall be supplied from a source specified as Electrical Energy Source Class 1
(ES1) to IEC/UL 62368-1, or specified as SELV according to IEC/UL 60950-1.
Voltage
12VDC (+/- 10%)
Current
1A
3.2
Connection Specification
The ICU device has two sockets on the back of the unit; a power connection socket
and a communication socket. The location of these is shown in
The drawing for the power cable is shown in
The communication cable is a standard USB Type-A to Type-B Micro USB cable. This
micro USB cable provides the physical link through which communications and video
are transmitted.
The maximum length of this data cable is 3 metres.
3.3
Screw Specification
The table below shows the screw specifications for the different screws on the ICU
device. The bezel screws are provided with the unit as it is already attached, the other
screws must be sourced separately.
Screw Position
Thread
Max Depth
Machine Mount Screw
M4
25mm
Bezel Screw
M4 (3mm Allen Key Head)
12mm
Camera Mount
Standard ¼” Tripod Mount
25mm
3.4
Bezel Mounting
It is possible to remove the front face of the ICU device to mount it inside a machine.
This lets you secure the face of the unit to the outside of the machine, while mounting
the ICU device behind it inside the machine:
ICU Device
Bezel