INNOVATEST DSAS Скачать руководство пользователя страница 4

1.

OPERATION

a.

The test specimen should be at least 6mm (0.25”) in thickness for the Shore A models and
at least 3mm (0.12”) in thickness for the Shore D models.

b.

The lateral dimensions of the specimen should be sufficient to permit measurements at least
12mm (0.5”) from any edge.

c.

The surface of the specimen should be flat and parallel over a sufficient area to permit 
thepresser foot to contact the specimen over an area having a radius of at least 6mm
(0.25”)from the indentor point.

d.

A specimen may be composed of plyed pieces to obtain the necessary thickness, but
determinations made on such specimens may not agree with those made on solid 
specimens because the surface faces between plyes may not be in complete contact

1.1

Testing procedure

a.

Ensure the tester is set to zero before commencing any tests.

b.

Change to REF I mode when using the durometer.

c.

Place the specimen on a hard, horizontal surface.

d.

Hold the durometer in a vertical position with the point of the indentor a least 12mm (0.5”)
from any edge.

e. 

Apply the presser foot to the specimen as rapidly as possible, without shock, keeping the
foot parallel to the surface of the specimen. Apply just sufficient pressure to obtain firm 
contactbetween presser foot and specimen.

f.

The hardness should be read off 1 second after the presser foot of the durometer has made
contact.

g.

In case of the specimen exhibiting a marked flow propensity, the reading may also be
taken after 15 seconds.

1.2

Using the supplied aluminium reference block

To check the tester has not been damaged simply place the 
indentor of the tester in the slot of the test block. Apply gentle 
pressure keeping the tester vertical, then simply read off the value 
obtained. This should be within ±2 units.

Please note:
INNOVATEST Europe BV recommend that the unit is returned
annually for re-calibration.

2

Manual DSAS/DSDS

Factory set 
calibration
screw 
Important do
not adjust

Содержание DSAS

Страница 1: ...Manual DSAS DSDS Digital Shore tester ...

Страница 2: ...ions can emerge due to new technologies and continuous development We reserve the right to change or modify specifications of products without prior notice We recommend you to contact our sales office for up to date information All rights reserved ...

Страница 3: ... Using the supplied aluminium reference block 2 2 The display 3 2 1 Set mode 4 2 2 Reference mode 4 2 3 Freeze display 4 2 4 Preset mode 5 3 Battery changing 6 4 Data transmission 6 5 IP65 protection 6 6 Warranty 6 7 Bench stand 7 Manual DSAS DSDS 1 ...

Страница 4: ...rometer c Place the specimen on a hard horizontal surface d Hold the durometer in a vertical position with the point of the indentor a least 12mm 0 5 from any edge e Apply the presser foot to the specimen as rapidly as possible without shock keeping the foot parallel to the surface of the specimen Apply just sufficient pressure to obtain firm contactbetween presser foot and specimen f The hardness...

Страница 5: ...s displayed Press SET briefly to activate electronics To switch off press SET and hold Automatic sleep condition occurs after 4 hours display off press MODE or SET to wake up electronics There are three electronic functions available SET REF I or PRESET To change between each mode press MODE and hold Manual DSAS DSDS 3 ...

Страница 6: ...libration value is entered in the preset pressing SET will display this value 2 2 Reference mode Press and hold MODE until REF I is displayed 2 3 Freeze display To freeze the display press SET briefly and HOLD will appear To unfreeze the display press SET briefly again 4 Manual DSAS DSDS ...

Страница 7: ...2 4 Preset mode Press and hold MODE until PRESET appears Press MODE to move cursor along Press SET to enter value or zero digit Press MODE and hold to store value and return to SET Manual DSAS DSDS 5 ...

Страница 8: ...et of the instrument 4 DATA TRANSMISSION Remove battery to connect Power Lead Opto cable Pressing SET sends a data transmission 5 IP65 PROTECTION Please note the IP65 protection rating for the instrument contained in these instructions applies to the electronics only 6 WARRANTY Except for battery exchange only the manufacturer has the ability to dismount the electronics Any tampering will invalida...

Страница 9: ...kg mass which is added to the 1kg mass to produce 5kg for Shore D Flip off the plastic dust cover from the top of the tester and screw in the adaptor with an M8 thread into the tester Place the test piece on the anvil of the bench stand and apply pressure to the operating lever so as to lift the anvil and test piece into the indentor of the tester Continue to apply pressure to the operating handle...

Страница 10: ...8 Manual DSAS DSDS ...

Страница 11: ...Manual DSAS DSDS 9 ...

Страница 12: ...705 Mobil 36 30 9349 611 Fax 36 1 424 1704 Web http www sidex hu E mail info sidex hu CORPORATE HEAD OFFICE MANUFACTURING DISTRIBUTION SERVICE INNOVATEST Europe BV Borgharenweg 140 6222 AA Maastricht The Netherlands Phone 31 43 3520060 Fax 31 43 3631168 Email info innovatest europe com www innovatest europe com IN09 056 V090712 ...

Отзывы: