Concentrating Pipette Select Quick Start Guide 

 (Patents and Patents Pending: www.innovaprep.com/patents) 

Page 

2

 of 

2

 

Installation of the Instrument Components: 

1.

 

Plug the 12V DC power supply into an available wall receptacle (110 -240 V AC 50/60 Hz 1.0 Amp) and then into the 

12V 

DC Jack

 on the side of the instrument.

 NOTE: Only use the power supply provided with the instrument. Use of any other 

power supply, including those from the previous Concentrating Pipette (CP-150), may damage the instrument and will 

void the warranty.

 

2.

 

Attach the 

Permeate Line

 to the 

Permeate Line Port

 using the quick connect fitting and insert the other end of the 

permeate tube into a waste container and secure using the supplied 

Permeate Line Clip

3.

 

Insert the provided 

Reusable Storage Fluid Adapter

 into a bottle of 

Storage Fluid

 and stow in the 

Storage Fluid Bracket

 

on the right side of the instrument. The 

Reusable Maintenance Tip

 is also stowed using this 

Bracket

The Start-Up Procedure: 

This procedure should be used the first time a new instrument is powered on

 

and each time the instrument is powered up after 

previously going through the Shut-Down procedure (the Shut-Down Procedure is described below). 

1.

 

Flip the 

Power Switch

 to turn the instrument on – it will take approximately 5 seconds for the unit to power up. 

2.

 

Press Enter to advance from the Welcome Screen to the Main Menu. 

3.

 

Select 

Maintenance>Start-up

 

4.

 

Follow the menu prompts to be guided through the Start-Up Procedure. 

5.

 

The instrument will now return you to the Run Protocol screen and is now ready for use. 

Selecting a Protocol: 

1.

 

On the Run Protocol screen the selected Protocol is displayed in all-caps above Start Run. 

2.

 

If a different Protocol is desired use the Arrow and Enter keys to select the displayed Protocol.  This will take you to the 

Protocol Setup screen.  

3.

 

Select 

Select Protocol

 and then use the Arrow keys to select the desired Protocol. 

4.

 

You will now be returned to the Run Protocol screen and the newly selected Protocol name will be displayed. 

Performing a Sample Concentration Run: 

1.

 

Select the CPT of your choice (See Table 2.1 Concentrating Pipette Tip Selection Guide in the User Guide) and insert the 

CPT into the Tip Interface Port. 

2.

 

Check that the Protocol listed on the Run Protocol screen matches the CPT type you are using. If not, follow the 

procedure provided above for Selecting a Protocol. 

3.

 

Place your sample container on the sample platform and lower the CPT into the sample until it contacts the bottom of 

the container. Select 

Start Run

. The vacuum pump will start to draw the sample through the CPT. When the entire 

sample has been processed, the instrument will conclude the run automatically. The menu will display the Run 

Complete screen which shows the processing time, and gives the option to Elute, Wash, or Return to the Main Menu. 

4.

 

Select 

Elute

. Follow the menu prompts to be guided through the Elute Procedure. Additional Elutions may be performed 

by selecting "Elute Again". 

5.

 

Remove the used CPT by pushing your thumb up against the bottom of the 

Fluidics Head

 while pulling the CPT down 

with your hand. 

Creating a Custom Protocol: 

1.

 

From the Select Protocol menu select FLAT (or the current protocol). This will advance you to the Protocol Setup menu. 

2.

 

Select 

Create Protocol

. You will be prompted to create a name for your protocol. Use the Arrow Keys to cycle through 

numerical and alpha characters sequentially from 1-9 followed by A-Z. To select a character, press Enter. Advance to the 

next menu by pressing and holding Enter. 

3.

 

You will be taken to a menu with the created protocol name at the top and standard or advanced settings for the 

protocol below. Standard settings will be shown if 

Protocol: Std

 has been previously selected in the System Settings and 

Advanced settings will be shown if 

Protocol: Adv

 was selected. System Settings may be accessed through the Main 

Menu. Descriptions of the standard and advanced settings are provided in Section 8 of the User Guide.

 

The Shut-Down Procedure: 

This procedure should be used at the end of each day of sample concentration runs. 

1.

 

From the Main Menu select 

Maintenance>Start-up

 

2.

 

Follow the menu prompts to be guided through the Shut-Down Procedure. 