Concentrating Pipette Select Quick Start Guide 

 (Patents and Patents Pending: www.innovaprep.com/patents) 

Page 

1

 of 

2

 

Instrument Overview: 

 

The 

Concentrating Pipette Select

 or 

CP Select

 uses 

Single-use Concentrating Pipette Tips (CPT) 

and

 Wet Foam Elution 

Canisters to 

allow users to quickly and easily concentrate biological particles from large liquid volumes (up to 5 L) into 

volumes as small as 150 µL. 

 

The 

CP Select

 allows the user to store up to 15 preset protocols. 

 

Factory set protocols are provided for Flat CPTs, Hollow Fiber CPTs, and Ultrafiltration CPTs. Using these protocols is a 

good place to start. 

 

Standard and Advanced Set Points are provided. Standard Set Points will provide the needed performance for nearly all 

applications and should be explored fully first. 

 

Start Up and Shut Down Maintenance Procedures are used with a Storage Fluid to maintain performance of the 

instrument, and should be used each day that the instrument is used. 

 

Important notes: 

 

The instrument can concentrate and elute with the 

Fluidics Head

 in any position. 

 

Never run alcohol stronger than 15% through the instrument as it may damage internal components. 

 

Use the 

Arm Tension Knob

 on the left side of the arm to adjust how tightly the 

Fluidics Head

 is held in place. 

 

Use the Up (

) and Down (

) arrows to highlight options on the 

Control Panel

 and press Enter (

 ) to select. To adjust 

settings, press enter and use the up and down keys to select the value desired. Then press the enter key again. At the 

bottom of each menu tree, the option “Return” is displayed; selecting this option will return you to the previous menu. 

 

The final volume of your sample is controlled most easily by the elution fluid Valve Open Time. The longer the open time 

and the more pulses performed, the larger the elution volume. 

 

Be sure that the 

CPT

 is firmly “clicked” into position; it will snap into the detent indicating that it is fully seated. 

 

Once a 

CPT

 has been used to concentrate a sample it cannot be reused. 

 

Protect the ‘Can Interface Port’ from contaminants by using the orange Elution Fluid Port Cover supplied with the 

instrument or by keeping an Elution Canister or the Storage Fluid Adapter in place. The port should be wiped with 

alcohol when changing fluids. 

 

Please keep the custom foam packaging insert for use in the event a return shipment is necessary.  

 

Prior to transporting or shipping the instrument refer to Section 6 of the User Guide for instructions. 

 

Refer to the chemical compatibility chart in the User Guide. 

 