Innova ALPHAiX Series Скачать руководство пользователя страница 2

1.

Dose Rate Measurement with Type G Counter Tube

The Type G end window counter tube is a pure contamination counter tube and not suitable for 
dose rate measurement. Simply because of its very high sensitivity, this counter tube becomes 
saturated relatively fast in cases of intense radiation which often arise in dose rate 
measurements. Its longer dead time leads to coincidence losses at a higher impulse density. 
Moreover, the all round radiation detection of the counter tube is not symmetrical due to the 
relatively large end window, so di

ff

erent results would be obtained at di

ff

erent angles to a 

specific radiation source. For dose rate measurement with ALPHAiX we recommend the Type A 
or B counter tube. The fluctuations in background radiation can be calculated in millirem per 
year (mrem/a) by multiplying the impulses by a factor of 4 and converting to 1 minute.

2.

Statistical Error of Measurement

120 mrem/a or 1.2 mSv/a is the usual background radiation (solar and earth radiation) which 
can, however, show considerable local variation. The usual background radiation of a location 
or test location can be determined if the radiation detection instrument is left running for 2 
hours with no radioactive radiation source present in its vicinity. The recorded number of 
impulses is converted to a value per minute. This value (Ipm) is then the so-called background 
count. In measurement, only measured values above the background count indicate the 
presence of radioactive exposure.

All measurements are subject to a statistical error of measurement. This is due to the fact that 
radioactive radiation does not occur constantly in time and space but at varying intervals.

The error of measurement is calculated from the root of the counted impulses:

Error of measurement in % = 

(N = total counted impulses)

This means the error of measurement declines as the number of impulses increases. In other 
words, the longer the measurement, the more accurate the measurement. Thus a series of 
measurements of, say, 100 impulses has an error of measurement of 10% but only 3.2% at 
1,000 impulses and 1% at 10,000 impulses.

For food inspections a minimum measurement duration of 10 minutes is recommended. 
Experience shows that the tolerance value for 10 minute measurements at background counts 
of 30 Ipm is at 35 Ipm (30 + 5), i.e. only the impulses in excess of 35 per minute are the result 
of additional radiation exposure. If a series of measurements is very close to the tolerance 
value, the series must be repeated for a longer measuring period.

OPERATING INSTRUCTIONS 

for Counter Tube Type G

JS Monday 10 January 2011                                                                                                                                                  Page  2 of  9

Содержание ALPHAiX Series

Страница 1: ...fferentiation of radiation types Page 6 7 Some more theory Page 7 APPENDIX Measuring table Page 8 Using the measuring table Page 9 ALPHAiX with Counter Tube Type G End window counter tube for measurem...

Страница 2: ...active radiation source present in its vicinity The recorded number of impulses is converted to a value per minute This value Ipm is then the so called background count In measurement only measured va...

Страница 3: ...to contact For accurate measurements a stand is required to ensure a constant distance for 10 minutes As already mentioned the tolerance value for a 10 minute measurement is 35 Ipm i e if a maximum of...

Страница 4: ...ust be extended if a 10 minute measurement does not show satisfactory results As you can see we have rounded up in the calculation of the tolerance value Needless to say you can work with the more acc...

Страница 5: ...with opened covers both at the tablet and the counter tube with the end window of the tube held directly on the tablet surface In the counter tube A the metal bezel can be put on the tablet as the en...

Страница 6: ...the two measurements If ALPHA radiation is present the first measurement without transparent film must be correspondingly higher The distance for these measurements should be 5 mm Separating the BETA...

Страница 7: ...tivity Bq of the radiation source required to allow its measurement The radiation energy keV and its activity Bq are two different factors which together with the type of radiation ALPHA BETA and GAMM...

Страница 8: ...n for this measurement Smaller distances give higher impulse frequencies and larger distances correspondingly lower counting efficiencies Impulses per minute Ipm NUCLIDE END WINDOW COUNTER TUBES IMMER...

Страница 9: ...ence shows that the pre requisites often do not concur with those of the measuring table A shorter distance usually 5 mm is often chosen for surface measurements using the type A or G end window count...

Отзывы: