Innomed Medical
Heart Screen 112C-1
Service manual
4/47
D-7443-1-C
3
Devices’ technical parameters
Classification according to 93/42/EEC directive:
II.a
This product has CE mark according to the 93/42/EEC Council
Directive on medical devices, dated on 14. June 1993.
3.1 General features
Reference temperature range
0... +40
C
Operating temperature range
Humidity
0... +45
C
20…95% humidity (without condensation)
Shipping/storage temperature range
-20... +55
C
Operating environment
dry, closed space
Operating and storage pressure
550 – 1060 hPa
Environment abstinence
IP20, the device protected from middle-sized solid objects
(>12mm) but it is not protected from water intrusion
Mains and power supply
Mains voltage
110-230VAC
Mains frequency
50 - 60 Hz
Mains power consumption
2A@110VAC 1 A@230VAC
Accumulator type
2100 mAh,
only Innomed produced accumulator is useable.
Accumulator nominal voltage
9,6 V
Accumulator charging time (informative)
In case of totally dead accumulator max. 2 hours
Accumulator charger type, protection
Intelligent quick charger, thermal protection at 54 C, thermal fuse at
90 C
Operating time from accumulator
60 minutes of ECG curve printing in intermittent operating mode, or
8 hours of monitoring (only basic functions are operated)
Safety
Safety class
I.
Patient protection (with PC connection as
well)
Defibrillator protected CF
Operating modes
Monitoring, display
3, 6 or 12-channel moni rhythm channel
Manual printing
Manual printing from any lead groups on 3, 6 or 12 channels
Automatic printing:
Adjustable length automatic printing from the 12 standard ECG leads in 2x6
or 4x3 channel groups.
Record storage (STORE)
Automatic recording and storage from 12 leads. Rhythm curve may be stored
in pre-programmed mode.
Analysis and diagnosis creation
(DGN)
Automatic 12-channel recording with analysis and diagnosis. Raw curve,
rhythm curve may be stored in pre-programmed mode. Data, marked average
curved appearance on the display and printer.
Arrhythmia analysis (ARTM)
Arrhythmia event storage, trends, statistics.
Result appearance on the printer and display.
Содержание Hear Screen 112C-1
Страница 14: ...Innomed Medical Heart Screen 112C 1 Service manual 14 47 D 7443 1 C 4 6 2 Devices base exploded diagram...
Страница 31: ...Innomed Medical Heart Screen 112C 1 Service manual 31 47 D 7443 1 C BOTTOM...
Страница 32: ...Innomed Medical Heart Screen 112C 1 Service manual 32 47 D 7443 1 C 9 2 HeartScreen112C 1 Interface Board TOP...
Страница 33: ...Innomed Medical Heart Screen 112C 1 Service manual 33 47 D 7443 1 C BOTTOM...
Страница 35: ...Innomed Medical Heart Screen 112C 1 Service manual 35 47 D 7443 1 C...
Страница 36: ...Innomed Medical Heart Screen 112C 1 Service manual 36 47 D 7443 1 C...
Страница 37: ...Innomed Medical Heart Screen 112C 1 Service manual 37 47 D 7443 1 C...
Страница 38: ...Innomed Medical Heart Screen 112C 1 Service manual 38 47 D 7443 1 C...
Страница 39: ...Innomed Medical Heart Screen 112C 1 Service manual 39 47 D 7443 1 C...
Страница 40: ...Innomed Medical Heart Screen 112C 1 Service manual 40 47 D 7443 1 C...
Страница 41: ...Innomed Medical Heart Screen 112C 1 Service manual 41 47 D 7443 1 C...
Страница 42: ...Innomed Medical Heart Screen 112C 1 Service manual 42 47 D 7443 1 C...
Страница 43: ...Innomed Medical Heart Screen 112C 1 Service manual 43 47 D 7443 1 C...
Страница 44: ...Innomed Medical Heart Screen 112C 1 Service manual 44 47 D 7443 1 C 10 2 HeartScreen112C 1 Interface Board...
Страница 45: ...Innomed Medical Heart Screen 112C 1 Service manual 45 47 D 7443 1 C...
Страница 46: ...Innomed Medical Heart Screen 112C 1 Service manual 46 47 D 7443 1 C...
Страница 47: ...Innomed Medical Heart Screen 112C 1 Service manual 47 47 D 7443 1 C 10 3 HeartScreen112C 1 Battery package...