View900 Cable Antenna Analyzer User Manual
INNO Instrument Inc.
Page 37 of 69
4.2.1.1.
OSL
C
ALIBRATION
Perform Open-Short-Load calibration using O-S-L Mechanical calibrator.
1.
Open Calibration
1)
Open cal start
2)
Open Cal Process
3)
Open Cal Complete
2.
Short Calibration
1)
Short cal start
2)
Short Cal Process
3)
Short Cal Complete
3.
Load Calibration
1)
Load cal start
2)
Load Cal Process
3)
Load Cal Complete
When calibration done successfully, the screen automatically returns to the measurement screen. CAL “On.”
Icon will be displayed.