7-4 Full Assembly Diagram
GRAPHIC FOR ILLUSTRATION PURPOSES ONLY
•
The technology should be used to measure only one subject’s
temperature at a time.
•
Measurements should not be solely, or primarily, relied upon to
diagnose or exclude a diagnosis of any disease.
•
Elevated body temperature should be confirmed with
secondary evaluation methods (e.g., an NCIT or clinical grade
contact thermometer).
•
Signage should instruct for the removal of glasses, headwear,
and masks.
•
Read the current government guidance regarding the use of
telethermographic systems for skin temperature measurements.
Additional information can be found by reading IEC 80601-2-
59:2017 Medical electrical equipment — Part 2-59: Particular
requirements for the basic safety and essential performance
of screening thermographs for human febrile temperature
screening and ISO/TR 13154:2017 Medical electrical equipment
- Deployment, implementation and operational guidelines for
identifying febrile humans using a screening thermograph.