background image

• 35 •

Storage of your Finetest™ Lite System

To prevent the meter and test strips from getting dirty, dusty or other contamination, please 

wash and dry your hands thoroughly before use.

Meter Storage

• 

Storage condition: 2~30

 or 36~86

 (Temperature) 

• 

Always store and transport the meter in its original storage case.

• 

Avoid dropping and strong impact.

• 

Avoid direct sunlight and humidity.

Test Strip Storage

• 

Storage condition: 2~30 

 or 36~86 

 (Temperature)

• 

Store your test strips in their original container only. Do not store the test strips in other 

containers.

• 

Store test strip packages in a cool and dry place. Keep away from direct sunlight and heat.

• 

Do not freeze.

• 

After removing a test strip from the container, immediately close it tightly.

• 

Record the discard date (date container opened, plus 6 months) on the test strip label.

• 

Do not use if it is expired(2 years) after manufacturing.

• 

Touch the test strip with clean and dry hands.

• 

Do not use if 6 months have been passed, since the opening of the container cap for test 

strips.

Control Solution Storage

• 

Storage condition: Store in a cool, dry place at between 8~30 

 or 46~86 

 

(Temperature)

• 

After use, close cap tightly.

• 

Make sure the control solution bottle is tightly closed.

• 

Record the discard date (date container opened, plus 3 months) on the control solution on 

the bottle.

• 

Do not use if expiry date has passed.

• 

It expires in 24 months after manufacturing date.

• 

It expires in 3 months after opening the container cap.

CAUTION 

1. Store the 

Finetest™ Lite test strip container in a cool, dry place. Keep out of direct 

sunlight. Do not freeze.

2. Store test strips in their original container only. Do not mix the test strips in new 

containers or in any other container.

3. Immediately replace the container cap and close tightly after removing any test 

strips from the container.

4. Make a note of the discard date, which is six months from the date you first open a 

new container of test strips. Throw Finetest™ Lite test strips and container away 

after the discard date.

5. Do not use the test strips after the expiration date printed on the package or 

container, Testing with expired strips it may cause inaccurate results.

6. Do not test at temperatures below 10

(50°F) or above 40

(104°F).

7. Do not test with humidity below 10% or above 90%.

8. Do not bend, cut, or alter the test strip.

9. Avoid getting dirt, food, and water on the test strip with wet hands.

10. Avoid getting dirt, food, and water on the color-coding label (backside of test strip).

Содержание Finetest Lite

Страница 1: ...Blood Glucose Monitoring System for self testing Instructions for Use 0 5ul 5 sec Auto Coding Auto Ejector...

Страница 2: ......

Страница 3: ...Blood Glucose Monitoring System for self testing Instructions for Use 0 5ul 5 sec Auto Coding Auto Ejector...

Страница 4: ...sional before making any changes to your diabetes management The Finetest Lite Blood Glucose Monitoring System is for in vitro diagnostic use only Important Information The Finetest Lite Blood Glucose...

Страница 5: ...n measures 2 Always use a new sterile lancet Lancets are for single use only 3 Avoid getting hand lotions oil dirt or debris on the meter lancets or lancet device 4 Please refer to page 33 for cleanin...

Страница 6: ...est with your finger tip 22 26 Test Strip ejector function 27 4 Reviewing Your Results Reviewing your blood sugar 28 Reviewing your averages 28 5 Alternate Site Test Lancing and sampling from an alter...

Страница 7: ...Icodextrin does not interfere with Finetest Lite test strips Do not use during or soon after xylose absorption testing Xylose in the blood will cause interference Before using any product please read...

Страница 8: ...tents If you find your seal has been broken please return it to the place of purchase 1 Finetest Lite Meter 2 Finetest Lite Test strip 3 Reusable Lancing Device 4 Lancets 5 Operation Manual 6 Warranty...

Страница 9: ...s test strip Serial Number Battery Cover Finetest Lite Meter CAUTION 1 Do not use Finetest Lite glucose test meter in a dry environment especially if synthetic materials are present Synthetic clothes...

Страница 10: ...rm Symbol Troubleshooting Symbol Average Test Result System Before Having a Meal After Having a Meal After Taking Medication After Sport Activity Control Solution Symbol Blood Drop Symbol for Test Tes...

Страница 11: ...test strips and container away after the expiry date and the container away after the expiry date has passed 5 Do not use the test strips after the expiration date printed on the package or container...

Страница 12: ...od glucose test strip 3 Practice using the lancing device and become accustomed with its use 4 Lancets and blood glucose test strips should be single use only 5 Wash your hands in warm clean running w...

Страница 13: ...ve daily alarms Power On Press for 3 seconds at least Date Time After pressing and releasing either or buttons to set the date and time respectively push button User Activity Option After pressing and...

Страница 14: ...n the alarm on off push button Temperature Unit After pressing and releasing either or buttons to set or and push button Alarm time After pressing and releasing either or buttons to set the alarm hour...

Страница 15: ...glucose test meter every time Check the code onon the container to see if the codes match CAUTION 1 Do not bend the glucose test strip to prevent the automatic code recognition failure 2 If the code...

Страница 16: ...lood glucose test meter or blood glucose test strip is not functioning properly If your blood glucose test results are not consistent with your symptoms or if you think results are not accurate If you...

Страница 17: ...so as not to be confused with the medication container This will allow you to understand differences between a control solution test and an actual blood test Fo If y yo Ple 4 Dip the test strip into t...

Страница 18: ...Do not use beyond expiration date Once opened the control solution expires after 3 months If your control solution test falls out of range please follow the next steps before contacting customer suppo...

Страница 19: ...glucose test and please contact Neon Diagnostics Ltd 2 The glucose control solution s measurement range has nothing to do with individual s blood glucose level as its own purpose is to check the statu...

Страница 20: ...s also comes with a transparent tip for AST Alternative Site Testing CAUTION 1 Never use a lancet that has been used by someone else Sharing lancets could lead to contamination 2 If the lancing device...

Страница 21: ...rom sunlight or moisture CLEANING AND DISINFECTION Clean and wipe the outer part of lancing device once a week at the minimum with a soft cloth dampened with mild soap and water Wash the adjustable ti...

Страница 22: ...he screen Er5 message will appear on the screen Please refer to page 40 Step 1 Firmly insert the test strip into the meter test port in the direction of arrow on test strip arrow down Please do not in...

Страница 23: ...single use only and the lancing device 3 Wash your hands thoroughly with soap and clean water after handling the meter lancing device and or test strip 1 Unscrew the lancing device tip 2 Insert a ster...

Страница 24: ...or thick or calloused skin 6 Pull the end of the lancet device back 7 To prick your finger push the center button CAUTION 1 If the blood smears or runs do not use this blood sample Dry the area and ge...

Страница 25: ...the blood downwardly it may cause device errors and failure Finetest Lite Test Meter only requires 0 5 ul sample volume Stick the edge of the test strip to the drop of blood The blood will automatical...

Страница 26: ...lect to use User Activity option during set up press the power button CAUTION 1 If the test result is out of the test range HI Lo message will be shown on the LCD window 2 Safely discard used glucose...

Страница 27: ...s waste in your area Please ensure you follow local regulations 1 Push the needle into the protective cover 2 Pull the lancet out and discard accordingly CAUTION 1 Pushing the ejector button too hard...

Страница 28: ...meter to calculate Press and release button The most recent result appears first Note the time and day of your blood sugar result Press and release the button and your previous result will appear on t...

Страница 29: ...o use your fingertips less often You may find that obtaining a blood sample from an alternative site is less painful than using a fingertip LIMITATION 1 Alternative site testing results should never b...

Страница 30: ...ly a heating pad to the site to increase blood flow Choose a fleshy area on the palm below your thumb or pinky finger Select a spot without any visible veins and away from any deep lines which may cau...

Страница 31: ...Insert lancet and replace with the clear tip 3 To bring fresh blood to the surface of the test site rub the test site vigorously for a few seconds until you feel it getting warm Applying heat may be h...

Страница 32: ...e spot may cause soreness and calluses 3 If bruising occurs at an alternate site or you have difficulty getting a sample consider sampling from a fingertip instead You may want to review the choice of...

Страница 33: ...hold bleach and water in 1 bleach 9 water ratio Dampen a cloth with this solution and wipe the lancing device thoroughly Soak only the tip for at least 30 minutes in the disinfectant solution Do not s...

Страница 34: ...light and heat Do not freeze After removing a test strip from the container immediately close it tightly Record the discard date date container opened plus 6 months on the test strip label Do not use...

Страница 35: ...h is six months from the date you first open a new container of test strips Throw Finetest Lite test strips and container away after the discard date 5 Do not use the test strips after the expiration...

Страница 36: ...r test results from the Finetest Lite meter to a computer You will need to download the software from the following web site www neondiagnostics co uk and and obtain the USB cable from Neon Diagnostic...

Страница 37: ...ur local battery collection systems and in compliance with your local environmental laws and regulations 3 Caution risk of explosion if battery is replaced by an incorrect type Battery Installation Th...

Страница 38: ...ry collection systems and in compliance with your local environmental laws and regulations Batteries contain chemicals that if released may affect the environment and human health The crossed out whee...

Страница 39: ...attery again and set the blood glucose test meter If the problem persists please contact Neon Diagnostics Ltd Caused by either used or wet blood glucose test strip Please insert a new glucose test str...

Страница 40: ...problem persists The ambient tempera ture is less than 10 50 Place the meter at a temperature between 10 40 50 104 for more than 30 minutes and test again The ambient temperature is over 40 104 Place...

Страница 41: ...m persists contact Neon Diagnostics Ltd Please insert a new test strip and perform your test again Please insert a new test strip and perform your test again There are no readings stored in the meter...

Страница 42: ...have episodes of hypoglycemia low blood sugar This may require immediate treatment according to your healthcare professional s recommendations Although this result could be due to a test error it is...

Страница 43: ...ithout obligation to incorporate such changes into previously manufactured instruments 3 Infopia Co Ltd has no knowledge of the performance of the instrument when the test strip is altered or modified...

Страница 44: ...ture 10 40 50 104 Operating Humidity 10 90 Dimension HxDxT 83 1 X 56 2 X 17 7 1 mm Weight Power Source 3V Li Battery CR2032 X 1 43 1g Including Battery Battery life One year after purchasing Test Stri...

Страница 45: ...See Instructions for use For In Vitro Diagnostic Use Batch code Do not reuse Serial number Date of Manufacture Manufacturer Keep away from sunlight Waste electrical and electronic equipment Symbol Tem...

Страница 46: ...1 132 Anyangcheondong ro Dongan gu Anyang si Gyeonggi do Korea www infopia21 com INFOPIA Co Ltd Bd General Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 732 60 03 E mail mail obelis net O...

Отзывы: