![Infinity IT-9800 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/infinity/it-9800/it-9800_user-manual_2056295019.webp)
19
Fixing
Issues
Trouble Shooting
Problem
Possible Cause
No power.
No Display on
Remote, but chair
does respond
to commands/
buttons pressed.
When reclining,
you can hear a
squeaking noise.
1.
Make sure power switch is on, the line icon (–) should
be pushed in – Test chair, continue to next step if
needed.
2.
Check each end of the power cord (where the power cord
plugs into the chair, and where it plugs into the wall outlet)
and make sure each end is securely plugged in– Test chair,
continue to next step if needed.
3.
Test wall outlet with another appliance – Test chair, continue
to next step if needed.
4.
Change fuse near power switch (7 AMP)(Reference page 13)
- Test Chair.
1.
Turn off power switch, the circle icon (O) should be pushed in.
2.
Unplug the power cord (for 10 seconds)
3.
Plug back in the power cord.
4.
Turn back on power on the switch, the line icon (–) should
be pushed in– Test chair remote, continue to next step if
needed.
5.
Check the each end of the wire which connects the remote
control to the chair, you will need to open up the remote
control (5 screws) to check these connection inside the
remote control – Test Chair.
1.
Determine which recline function is making the squeaking
noise by testing each of the reclining motors (Footrest,
Backrest, and Rotate).
2.
Once you determine which motor/s are making the
squeaking noise, lubricate where the motor travels in and out
of its housing (underneath the chair) with either WD-40 or 3
in 1 Oil – Test Chair, continue to next step if needed.
3.
Use either WD-40 or 3 in 1 Oil to lubricate the side bolts
on the chair (which connects the backrest to the chair seat
frame). These bolts are located in between the chair’s
armrest and each side of the backrest – Test Chair.