background image

User Manual | 2.4mm Coaxial Calibration Kits 

 

 

4

 

   

 

 

 

 

 

 

 

 

       

pasternack.com

 

 

General Information 

Calibration Kit Description 

These Pasternack 2.4mm coaxial calibration kits are designed to provide accurate calibrations of network 
analyzers in the DC to 50.0 GHz range. These kits include all the necessary calibration standards and 
associated hardware needed for the accurate calibration of most network analyzers. This manual applies 
to the following model numbers: PE5CK1015 9-piece SOL (Short, Open, Load) calibration kit and 
PE5CK1016 12-piece SOLT (Short, Open, Load, Through) calibration kit.  

Refer to th

Calibration Kit Contents

  section  for information on included components and available kit 

options. 

NOTE: This document, along with the kit data file and datasheet, can be downloaded from pasternack.com. 

 

Maintenance 

This calibration kit is relatively maintenance-free, if the components are handled with the same care that is 
appropriate to all precision equipment. As with any precision component, proper care should be taken to 
ensure clean mating surfaces, correct alignment when mating, and proper torqueing of connectors. To help 
maintain the integrity of the components in the kit, routine visual inspection and cleaning of mating surfaces 
is recommended. Failure to do so may result in degraded repeatability and accuracy, as well as damage to 
any mated devices. 

 

Calibration 

To maintain and certify the calibration kit’s ongoing performance to specification, we recommend that all 
kits be periodically returned to Pasternack for calibration. The typical calibration cycle is one year, although 
actual needs may vary depending on usage.  

 

Supporting Test Port Adapters 

When configuring a test setup, ensure that damaging stresses are not applied to the connectors on the test 
set. This is particularly critical when the attached components are heavy or long. Always properly support 
the test port adapters being used. 

 

Electrostatic Discharge Precautions 

Protection against electrostatic discharge (ESD) is essential while inspecting, cleaning, or making 
connections to connectors attached to a static-sensitive circuit, such as those found inside test sets. 

When handling the connectors on the test set, be aware that you are coming into contact with exposed 
center conductors that are connected directly to the static-sensitive internal circuits of the network analyzer. 
Ensure that you and your equipment are well-grounded before inspecting, cleaning, or making connections 
to test set ports. Standard ESD precautions, such as the use of grounded wrist straps and grounded 
antistatic mats, are recommended.

 

 

Содержание PE PASTERNACK PE5CK1015

Страница 1: ...f Infinite Electronics Inc User Manual 2 4mm Coaxial Calibration Kit DC to 50 GHz Models PE5CK1015 PE5CK1016 Pasternack PO Box 16759 Irvine CA 92623 Phone 866 727 8376 or 949 261 1920 Fax 949 261 7451...

Страница 2: ...t be returned to Pasternack and must be issued a return authorization number by Pasternack prior to shipment Buyer shall prepay shipping charges to Pasternack Obligation is limited to the original Buy...

Страница 3: ...2 General Information 4 Calibration Kit Description 4 Maintenance 4 Calibration 4 Supporting Test Port Adapters 4 Electrostatic Discharge Precautions 4 Connector Description 5 Connector Care 5 Connect...

Страница 4: ...tion and cleaning of mating surfaces is recommended Failure to do so may result in degraded repeatability and accuracy as well as damage to any mated devices Calibration To maintain and certify the ca...

Страница 5: ...should be visually inspected under magnification and cleaned on a regular basis Proper connector contact pin depths should also be verified through regular inspections using a connector gage kit to en...

Страница 6: ...while torquing the nut on the other device or cable This open end wrench is supplied with the calibration kit for this purpose Using the torque wrench Hand tighten the connection being torqued by hold...

Страница 7: ...le PE5TR1006 1 ea Load male PE5TR1007 1 ea 1 4 X 5 16 Open End Wrench PE5TL1001 1 ea 7 16 X 1 2 Wrench PE5TL1002 1 ea 5 16 Torque Wrench PE5019 1A 1 ea Short female PE5SC3010 1 ea Short male PE5SC3011...

Страница 8: ...s made by the instrument The calibration standards in this kit have precisely known electrical behavior and during calibration the VNA software uses the raw measurement data and the known behavior of...

Страница 9: ...2019 Infinite Electronics Inc Pasternack is a registered trademark of Infinite Electronics Inc 9...

Страница 10: ...omponents keywords calibration ports vna sort y searchtype 1 view_type grid Pasternack Test and Measurement Products https www pasternack com nsearch aspx Category Calibration Kits Portable Calibratio...

Страница 11: ...Infinite Electronics Inc 11 Contacts Customer Support Sales Pasternack PO Box 16759 Irvine CA 92623 USA Phone 866 727 8376 949 261 1920 Fax 949 261 7451 Sales Email sales pasternack com Customer Supp...