10 – 18
Transpector SPS Operating Manual
10.6.2.2.2 Data Threshold
Defines the threshold that the
Start Mode Type
utilizes to determine when the run
will start. The
Data Threshold
is set in units corresponding to the
Start Mode Type
selected.
Maximum Signal
is in units of current,
Pressure Above/Below
is in
units of pressure, and
Analog 1 Above/Below
is in units of volts. (See
.)
Figure 10-29 Data Threshold
10.6.2.2.3 Emission
Defines the mode of the filament when the run starts. (See
If
Emission
is not turned on, data will not be collected.
Figure 10-30 Emission
As Is
uses the current status of the filament.
Off
starts the data acquisition with the filament off.
On
starts the data acquisition with the filament on.
10.6.2.2.4 Multiplier
Defines the mode of the electron multiplier when the run starts. (See
Figure 10-31 Multiplier
As Is
uses the current status of the EM.
Off
starts the data acquisition with the EM off.
On
starts the data acquisition with the EM on.
10.6.2.2.5 Relay 1
Defines the state of
Relay 1
when the run starts
As Is
leaves the relay in its current state at the beginning of the run.
Open
opens the relay at the beginning of the run.
Close
closes the relay at the beginning of the run.