7 - 6
Transpector MPS Operating Manual
7.3 Communication Problems
To correct communication problems, check the following:
Is the Ethernet cable connected to Transpector MPS and the host computer
(either directly or through a router/switch)?
Does the Transpector MPS IP address have the same network prefix as the
host computer?
Is Port 80 open on the host computer?
Is there an IP address conflict between Transpector MPS and another network
device?
See
Chapter 2, Connecting Transpector MPS
communications problems.
High noise
level
System grounding
Verify that vacuum system is
grounded.
Electronics failure
Return to INFICON for repair.
Transpector MPS
electronics box not mounted
properly on sensor
Push Transpector MPS
electronics box all the way on
to the sensor.
EM defective
Replace EM assembly or
sensor.
Scan speed too fast for gain
setting
Increase dwell time.
Table 7-1 Symptom—Cause—Remedy Chart (continued)
SYMPTOM
CAUSE
REMEDY