10 – 26
Transpector CPM Operating Manual
10.6.2.4.2 Dwell Time
Defines the amount of time in milliseconds (ms) that FabGuard Explorer will collect
data at every data point. In general, the longer the dwell time, the lower the noise
of the signal. Longer dwell times at each point will result in longer overall scan
times.
Dwell Time
is only selectable when
Bin Independent
is chosen as the
Dwell Mode
Figure 10-45 Dwell time
10.6.2.5 Delay
Defines whether FabGuard Explorer will add additional wait time after the sensor
has completed a full scan.
10.6.2.5.1 Delay Mode
Figure 10-46 Delay mode
Minimum
sets the
Time Between Scans
to its minimum value. After the
sensor has completed a scan, no additional wait time will be added to the scan
by FabGuard Explorer.
Manual
allows the operator to set an additional delay time after the sensor
completes the scan.
Содержание Transpector CPM
Страница 1: ...O P E R A T I N G M A N U A L PN 074 641 P1B Cover Page Transpector CPM Compact Process Monitor ...
Страница 3: ...Declaration Of Conformity Page 1 ...
Страница 74: ...5 12 Transpector CPM Operating Manual Figure 5 6 Air mass spectrum N N2 O2 Ar O H2O ...