10 – 25
Transpector CPM Operating Manual
10.6.2.3.6 Relay 1
Defines the state of
Relay 1
when the run stops.
As Is
leaves the relay in its current state at the end of the run.
Open
opens the relay at the end of the run.
Close
closes the relay at the end of the run.
10.6.2.4 Dwell
Determines how and for how long the RGA collects data at each point.
10.6.2.4.1 Dwell Mode
Determines how the dwell is chosen for each data point. This option is changeable
only when in
Selected Masses
mode. (See
Figure 10-44 Dwell mode
Bin Independent
uses the same dwell time for each data point.
Bin Specific
is only available when running in Selected Masses mode. This
choice allows for each mass to have a specific dwell time. This methodology is
desirable if a peak of interest has an unusually noisy response.
FabGuard Explorer can be set to dwell at that peak for a longer time period,
which will decrease the noise and increase the signal-to-noise ratio.