How to Set your IRwin Methane Leak Detector  

for 1 Point Calibration 2.2% (2.5%)

17

18

Use          to select 

Disabled

 and press

      

19

4 of 5

mimc66en1-02 (1905)

20

In 

Universal Low Range

, use        to select the  

second choice in the drop-down menu, press

 

     

Navigate to far right drop-down menu and press

      

Press         again to view the drop-down menu

In 

Universal Low Range

, use        to select the third 

choice in the drop-down menu, press

      

Содержание IRwin Series

Страница 1: ...et unit QUICK START GUIDE 1 of 5 1 Make sure your IRwin is set in Universal Mode see the How to set your IRwin for Universal Mode quick start guide Use to navigate to General and press 2 3 4 5 Press followed by Use to navigate to AbG Cal Setup and press mimc66en1 02 1905 Use to navigate to Custom and press ...

Страница 2: ...ollowed by 8 9 2 of 5 mimc66en1 02 1905 Use and to navigate to Calibration and press Use to navigate to Settings and press 10 If prompted enter the min Intermediate password code If not prompted go ahead to next step 6 If prompted enter the Supervisor Advanced password code If not prompted go ahead to next step ...

Страница 3: ...nt Calibration 2 2 2 5 3 of 5 mimc66en1 02 1905 Press to get to the drop down menu 12 Press on the first choice of the drop down menu 13 Navigate to far right drop down menu and press 14 Use to select Disabled and press 15 Press again to view drop down menu ...

Страница 4: ...Disabled and press 19 4 of 5 mimc66en1 02 1905 20 In Universal Low Range use to select the second choice in the drop down menu press Navigate to far right drop down menu and press Press again to view the drop down menu In Universal Low Range use to select the third choice in the drop down menu press ...

Страница 5: ...r respective owners 2019 INFICON mimc66en1 02 1905 How to Set your IRwin Methane Leak Detector for 1 Point Calibration 2 2 2 5 21 22 23 Select Set by using and and press Navigate to far right drop down menu and press Use to navigate to drop down menus Set the actual value of the calibration gas you have nominally 2 2 or 2 5 by entering 2 and press 2 5 and press 24 Press followed by and ...

Отзывы: