IC6 Operating Manual
10.4.34 Remote Layer Action
Command
= RL <Command ID> <Layer Number> (<Action Value>)
Command ID
Layer Number
= <Byte> 0 - 200
If Zero is given as the layer number then the remote action will be for the active
layer. If more then one layer an error will be returned.
Action Value
= <Byte> | <Integer> | <Float>
See
for commands that require an action value.
Response
= None (Just header and trailer)
36
(0x24)
RateWatcher
Sample Initiate Off
<Material >
<Byte>
Turns RateWatcher sampling off.
37
(0x25)
Clear Q and S
counts for sensor X
<Byte>
Clear Q and S counts for specified sensor X
38
(0x26)
Name Material
<Material><Library
Entry>
<Byte><Byte>
Give material number <Material> (1-32) the name
corresponding to material entry <Library Entry> (0-254).
Library entries range from 0 (Ag) to 254 (ZrO
2
), with
numbers correcponding to the alphabetical listing in the
library. The density and Z-Ratio will be changed
accordingly.
Table 10-15 Remote general (continued)
RG
Command
ID
Meaning
Action Value
Notes
Table 10-16 Remote layer
RL
Command
ID
Meaning
Action
Value
Notes
0 (0x00)
Suspend
Cannot be done from Idle
1 (0x01)
Zero Thickness
Cannot do from Suspend
2 (0x02)
Start Deposit
Will always zero thickness on entry.This cannot be done from
Ready, Idle,Stop, Suspend, Crystal Switching or Crucible
Switch. When there is no good crystal, this command will go
to "Time Power" if the previous state was Non-Deposit Hold,
otherwise, is not allowed.
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...