5xx Flow and pressure errors
W543 Flow in the sniffer line is too low
– Filter in the sniffer tip is clogged
– Sniffer tip is clogged or defective
– Replace the filters
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W544 Valve does not switch back and forth
Internal defect on the sniffer line
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W545 Flow through measurement line is too
low
– The filter in the sniffer tip is clogged.
– Sniffer tip is clogged or defective.
– Replace the filters
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service!
W546 Leak on the measurement line
– The filter in the sniffer tip is clogged.
– Leak or defect in the sniffer tip or the sniffer line.
– Replace the filters.
– Check the plug and screw connections.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W547 Flow through reference line is too low
– Filter in the sniffer tip is clogged.
– The sniffer tip is clogged or defective.
– Replace the filters.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W548 Leak on the reference line
– The filter in the sniffer tip is clogged.
– Leak or defect in the sniffer tip or the sniffer line.
– Replace the filters
– Check the plug and screw connections.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W549 Measurement and reference line are
mixed up
– Poor calibration.
– Internal defect on the sniffer line.
– Recalibrate the device.
– Contact INFICON customer service.
Table 10: Warning and error messages (Contin.)
No. Message
Possible
error
sources
Troubleshooting
Содержание HLD6000
Страница 6: ...6 Table of Contents Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 12: ...12 3 Shipment Transport Storage Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 22: ...22 4 Description Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 28: ...28 5 Installation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 54: ...54 6 Operation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 60: ...60 7 Warning and error messages Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 71: ...10 Appendix 71 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 72: ...72 10 Appendix Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 74: ...74 Index Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Страница 75: ...75 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...