8
about 1 hour of useful charge in the power cells. To recharge the
power cells, use one of the instrument’s power cords.
WARNING:
DO NOT ATTACH ANY OTHER POWER
CORDS TO THE D-TEK’S POWER JACK EXCEPT THE AC
POWER ADAPTER AND DC CORDS DESIGNED FOR THIS
INSTRUMENT.
• Connect the AC power adapter cord to the instrument, and plug
the adapter into a working outlet
or
• Connect one of the DC cords to the D-TEK, and tap the power
from your vehicle’s battery.
Allow 10 hours for a full charge. You can operate the leak detector
on any of its power cords while the power cells are charging. Using
the instrument while the power cells are charging, however, will
lengthen the time required for a complete charge. The leak
detector will not overcharge the power cells, but for maximum
power cell life, disconnect the power adapter when the power cells
are fully charged.
Maintaining Your INFICON D-TEK
Recharging the Power Cells (Continued)
Probe
Sensor
Filter
Tip protector
Changing the Filter
The foam filter at the probe tip should be replaced if it becomes
plugged with water or oil. To replace the filter, unscrew the tip