10 | Accessories and interfaces
INFICON
132 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
10.4.1.2 Setting the LAN connection in the PC or Tablet
LAN connection - quick start
If you have performed the steps described here once, it is sufficient to enter the IP
address for many devices in case of repetition.
ü
The PC is connected to the same network as the leak detector.
ü
A LAN setting has been set up in the configuration of the leak detector, see also
“Configure the LAN connection of the leak detector [
131]“.
ü
You have noted down the IP address of the leak detector. It can be found in the
leak detector under "Diagnosis > Information > Device > Network".
ü
Java Script is enabled in the web browser. We recommend using a current version
of the web browser Chrome™, Firefox® or Safari®.
► To access the leak tester from the PC or tablet web browser, enter the IP address
of the leak detector as follows:
http://<IP address>
ð
The current active user interface of the leak detector is called.
ð
The same functionality is available on the PC or Tablet as on the touchscreen of
the leak detector.
10.4.1.3 Allow client access
ü
Supervisor rights
1
> Setup > Network > Client access
2
To allow the leak detector to be operated by PC or Tablet, activate the "Client
Access" option.
ð
If the option "Client Access" is not activated, the leak detector cannot be
operated via PC or Tablet. No settings can then be changed.
3
Save
.
Display of all connected clients
ü
The network connection between the leak detector and one or more PCs or Tablets
has been established, see “Setting the LAN connection in the PC or Tablet [
►
> Setup > Network > Connected clients
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......