3000 Micro GC Operating Manual
4.2.4.5 Connecting the Carrier Gas to the 3000 Micro GC
1
Connect the carrier gases to the instrument via the 1/8 in. Swagelok fitting(s)
located on the back of the instrument. See
.
WARNING
If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.
CAUTION
Do not use leak detection fluids.
4.2.4.6 Using the Onboard Carrier Gas Cylinder (Portable 3000 Micro GC only)
For information on refilling the internal carrier gas cylinder, see
Refill Onboard Carrier Gas Cylinder, on page 8-43
.
1
Remove the shipping caps.
2
Ensure the carrier gas knob on the front of the instrument is set to
OFF
.
See
Figure 4-6 Carrier gas knob
3
Connect the
CARRIER OUT
port to the
CARRIER IN
port with the carrier
jumper tube as show in
CARRIER IN
port 1 if 3000 Micro GC
has two ports available.