3000 Micro GC Operating Manual
8.3.5 Running a High Pressure Gas Sample with Liquids/Particulates
1
If the Pressure Reducer and Genie Filter Assembly (PN G2816A) are not
installed, install them as instructed in
section 5.1.5, Installing the Pressure
Reducer and Genie Filter Assembly, on page 5-12
NOTE:
The Pressure Reducer and Genie Filter Assembly can connect directly
to 3000 Micro GC or to the sample inlet filter (PN G2801-60900).
NOTE:
There is no set range for the flow meter. Verify some flow is present.
Figure 8-32 Pressure reducer and Genie filter assembly sample conditioner (PN G2816A)
2
Turn 3000 Micro GC.
3
Check that the carrier gas supply is sufficient to run all of the samples.
4
If using a sample vessel, prepare the sample.
5
Connect the sample vessel or gas source to the NPT sample input fitting on the
Genie Filter Assembly.
6
Open the sample vessel valve.
7
Use EZ IQ to load the desired method and perform the analysis.
8
When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.
9
Repeat steps 4 to 8 to run the next sample.
Adjustment
knob
NPT sample
input fitting
To sample input
fitting on
3000 Micro GC
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...