8.1.2 Shutdown/Restart
CAUTION
Avoid shutting down 3000 Micro GC using the power
switch on the front panel. This may cause a system
integrity issue.
CAUTION
If a 3000 Micro GC is powered off when the column is at a
high temperature, column contamination and
degradation may result.
Proceed as follows to shutdown or restart a 3000 Micro GC.
1
Create a shutdown method:
Turn off the
Detector Filaments
Lower the
Column Temperature
to 45
°
C
Select
On
for
Column Pressure
and set to at least 20 PSI
A shutdown method is illustrated in
.
Figure 8-2 Example shutdown method
2
Download the method from
EZ IQ
onto the 3000 Micro GC.