The Pool eXact

®

 EZ photometer has a 5 minute automatic shutdown timer.

• 

Each test menu can store 20 results. To retrieve the stored results, go to the desired 

test using the MENU key. When the desired test is displayed, press and hold down the 

MENU key. Continue holding down the MENU key to scroll the stored results for that 

test, starting with the most recent result. The meter will display, from memory, the last 

20 readings in sequence beginning with -20, which is the latest result, followed by -19, 

which is the 2nd latest result, etc; and finally -01, which is the oldest result retained. 

Only the last 20 readings are stored in each menu. This meter is able to store 100 

results in memory (20 in each menu).

• 

Before testing, rinse the meter sample cell with the sample water 3 times.

• 

Pool water samples should be taken 18 inches below the surface. Submerge the meter 

with open cell facing down 18 inches then turn meter upright at that depth to fill the cell. 

Remove meter from water with the sample for testing.

• 

Always fill the cell to capacity (4ml); be careful not to splash liquid over the side.

• 

Test immediately after filling the cell with the water sample. 

• 

To obtain optimal accuracy when testing outdoors (sunlight), use the cell cover when 

zeroing and reading the sample.

• 

Due to the strip slitting process, you may find one or two strips that are noticeably 

smaller or larger in width than the normal strips in the bottle. These should be 

discarded. Using these strips may give unreliable results.

• 

Meter is not compatible for use with powder pillows, tablets, or liquids from other 

manufacturers.

• 

Dip strip for entire countdown.

• 

Each eXact

®

 Strip Micro is valid for ONLY one test. Discard strip after use.

• 

Dry the outside of the meter before storage.

• 

Remove batteries before storing for prolonged periods.

• 

Store the meter and test materials out of direct sunlight and away from chemical 

storage areas. 

• 

Minimize exposure of meter and test reagents to heat above 90°F (32°C).

• 

When installing batteries, verify the O-ring is still attached to the screw before 

tightening.

• 

Cleaning the cell with water and brush after each test is recommended for best 

accuracy and prior to storage of unit.

TIPS

Issue Cause 

Solution

Dim screen or no 

Low battery 

Replace batteries 

response from meter
 

Low battery 

Replace batteries 

LO on LCD Display 

Dirty cell 

Clean cell (see pg.15) 

while zeroing 

Cloudy sample 

Dilute sample or use sieve 

 

Bad LED 

Contact ITS.

HI on LCD Display 

Result above 

 

while reading 

detection level 

Re-run test to verify result

LO on LCD Display 

Result below 

 

while reading 

detection level 

Re-run test to verify result

LO flashes on LCD 

Improper test procedure 

Re-read test procedure 

Display, then Err 

for Combined Chlorine 

and follow directions carefully

HI flashes on LCD 

Combined Chlorine result  Re-run test to verify result 

Display, then Err 

above detection level 

Dilute and re-run test

Total Chlorine result 

Improper test procedure 

Re-read test procedure 

 

less than Free Chlorine   

and follow carefully

TrOUBleShOOTIng

Содержание Pool eXact EZ

Страница 1: ...1 QuicK stArt guiDe ADVANCED PHOTOMETER SYSTEM DigitAl AccurAcy For better pool cAre ...

Страница 2: ...rewdriver to remove the screw from the base of your Pool eXact EZ Photometer 2 Remove the base 3 Install 4 new AAA batteries as illustrated inside your Pool eXact EZ Photometer We recommend using high quality batteries 4 Replace the base firmly with pressure while tightening the screw The meter will power on automatically COnTenTS WelCOMe Welcome Kit Contents 2 Battery Installation 2 Meter Overvie...

Страница 3: ...er testing CELL COVER Covers the cell for mixing and bright light situations Combined Chlorine DPD 3 Test requires Free Chlorine DPD 1 486637 to be run first For resellers and distributors products sold in case quantities 12 units per case PARAMETER TEST No PART RANGE 1 2 3 4 5 6 7 For use in accordance with U S Patent No 7 333 194 U S Patent No 7 491 546 South African Patent No 2007 0628 EU Patent...

Страница 4: ...p Micro e g Free Chlorine DPD 1 from the bottle or foil packet and set in a dry convenient place Turn meter on by pressing ZERO ON The display will indicate current MENU followed by the last test result Press and re press MENU until the display shows the parameter to be tested e g CL1 matches CL on bottle STAnDArD STrIP MeThOD 1 2 3 FIll DIP reAD CL1 ...

Страница 5: ...ple cell and simultaneously press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes sec Remove and discard the strip after 1 on the display disappears The cursor will move across the display while the meter prepares to measure the sample reAD result displayed After testing is complete rinse the sample cell immediately ...

Страница 6: ...acket Set it in a dry convenient place With the DPD 1 sample still in the cell press ZERO ON The display reads 0 PPM Dip the eXact Strip Micro CL DPD 3 into the CELL and simultaneously press READ This starts another 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes sec Remove and discard the strip after 1 on the display disappears FIll DIP...

Страница 7: ...er measures the sample reAD the Combined Chlorine value displayed If you require a value for Total Chlorine press READ again and record result displayed After testing is complete rinse the sample cell immediately FIll DIP reAD 11 12 SCAn FOr TeST DeMO VIDeO ...

Страница 8: ...nd re press MENU until the display shows the parameter CY5 Rinse the sample cell 3 times with the water sample you will be testing and fill the sample cell Using the cell cover cap the sample cell and press ZERO ON The cursor will move across the display followed by 0 PPM 1 2 3 4 FIll DIP reAD CY5 CYAnUrIC ACID TeST ...

Страница 9: ...inger over cap and mix the sample by turning the meter upside down repetitively during the 20 second countdown When timer displays 1 place meter on flat surface upright and the meter begins a 60 second count up timing at the end of which the cursor will move across the display while the meter prepares to measure the sample Record result displayed this result is automatically stored in CY5 Rinse th...

Страница 10: ...eXact EZ photometer uses a 525nm wavelength The eXact Strip Micro CL DPD 1 uses the same reagents and proportions and the resulting solution pH is maintained between 6 2 and 6 5 as specified by AWWA method 4500 Cl G The USEPA does not approve commercial DPD delivery systems The eXact Strip Micro CL DPD 1 for Free Chlorine and the eXact Strip Micro CL DPD 3 for Combined Chlorine or the eXact Strip ...

Страница 11: ...in the bottle These should be discarded Using these strips may give unreliable results Meter is not compatible for use with powder pillows tablets or liquids from other manufacturers Dip strip for entire countdown Each eXact Strip Micro is valid for ONLY one test Discard strip after use Dry the outside of the meter before storage Remove batteries before storing for prolonged periods Store the mete...

Страница 12: ...e at http www sensafe com product registration form Personal data is kept confidential The eXact photometer is warranted to be free from defects in materials and workmanship for a period of two 2 years from the date of purchase by the customer ITS will repair or replace at its discretion product which is deemed to be faulty due to manufacturing defect Warranty does not cover product damage caused ...

Отзывы: