Cadmium in Water Test Procedure
Uses Reagent Set Part Number 486904
1
PREPARE SAMPLE FOR TESTING
Collect water sample in 50mL conical tube to the 50mL
line. Add five (5) drops of
HCl-1 Reagent, Part #486994.
Mix and allow to sit for at least five (5) minutes. After this
time, TEST SAMPLE is ready for testing.
2
TURN METER ON
Press the
button to power the meter on; the
display will show all annunciators, then the current MENU
selection, followed by the last reading.
4
ADD SAMPLE TO CELL
Using the TEST SAMPLE from above, fill and empty
the
CELL four (4) times. Finally, fill cell to capacity with
the TEST SAMPLE. Tilt meter forward to allow excess
sample to flow out in order to make room for Pb-2
Reagent addition below.
5
ADD REAGENT PB-2
Add five (5) drops of
eXact
®
Reagent Pb-2, Part #488375-B.
1
2
3
SELECT TEST: Cd
Press and re-press the
button until the display
shows the parameter Cd.
4
5
6
DIP STRIP & PRESS READ
Dip the
eXact
®
Strip Pb-3, Part No. 486996 into the
CELL and immediately press
. This starts the
20 SECOND countdown timer. During this time move
the strip in a gentle back and forth motion.
Remove and
discard the strip after “1” on the display disappears.
The display will flash (- - -) and begin immediately
counting up from
1 to 240. After the 240 seconds, the
meter automatically zeros. The cursor will move across
the display followed by
0 µg (µg/L).
7
DIP STRIP & PRESS READ
Dip the
eXact
®
Strip Pb-4, Part No. 486995 into the
CELL and immediately press
. This starts the
20 SECOND countdown timer. During this time move
the strip in a gentle back and forth motion.
Remove and
discard the strip after “1” on the display disappears.
The display will flash (- - -) begin immediately counting
up from
1 to 240. After the 240 seconds, the cursor will
move across the display, informing you that it is about
to measure the sample as
µg (µg/L). Record result
(this result is automatically stored in
PB). After testing
is complete discard sample and rinse cell at least three
times with clean water.
6
7
3