22
BLOCK DIAGRAM 3
PARKING CYCLE
Alarm 3 operates in the same way as the normal cycle.
The times defined for alarms 1 and 2 (time delays) are, in this case, used to define the minimum
operating time for each relay.
t AL1
tR1 R2
t AL2
Threshold AL2
Threshold AL1
Starting of
relay R1
Stopping
of relay R1
Starting of
relay R2
Stopping of
relay R2
min.
t
AL1
Min. operating time for alarm 1
(defined for each channel)
t
1
t
AL2
Min. operating time for alarm 2
(defined for each channel)
t
2
t
R1 R2
Switching time from relay 1 to relay 2
(defined for the whole unit)
t
R1 & R2