TECHNICAL BULLETIN 

 

10 July 2014

 

Issue 14-7.8 

Revised MX6 Refuge Chamber  

Maintenance Recommendations 

Industrial Scientific’s MX6 Refuge Chamber kit is an approved MSHA Part 7 Air Monitoring component. 
Please see below for revisions to the monthly maintenance procedures and two warnings to ensure your 
instrument is working properly.  
 

Revised Monthly Maintenance Procedure 

The previous recommendation instructed users to remove the alkaline battery pack and install a 
spare battery pack before performing monthly maintenance. The previous recommendation then 
stated the user should reinstall the alkaline battery pack once maintenance was complete as the pull 
tab should only be removed with a deployed refuge chamber in an emergency situation. 
 
The new recommendation is to remove the pull tab monthly and perform all maintenance steps 
using the alkaline battery pack. Three new alkaline batteries and a pull tab should be reinstalled 
prior to returning the instrument to the case. Please see the next page for an excerpt from the 
updated “Use and Service Instruction Manual” with additional details on the new MSHA-approved 
process. 
 

Caution – Over Tightening Instrument Screws 

Over tightening screws may permanently damage the instrument, particularly the battery pack. 
Please consider using a torque screwdriver with a 65 in-oz. specification. 
 

Caution - Sample Tubing Damaged during Kit Repackaging 

Sample tubing may be permanently damaged if compressed by the case lid if the contents are 
repackaged incorrectly. Once the tubing is damaged, the instrument’s pump may not properly 
deliver a sample to the sensors and cause the instrument to produce a pump fault. Please be careful 
to always repackage the case properly. 

 

Updated instruments manuals (Part# 17148420) and spare pull tabs are available (Part # 17155260). 

 

For additional information on the MX6 Refuge Chamber kits, contact us by email at 

[email protected]

 or by phone at 1-800-DETECTS (338-3287). 

 
Sincerely, 

 

Global Product Manager, Multi-Gas Solutions  

 

 