1
1705-8868
Main PC Board (1)
2
1705-6755
Display PC Board (1)
3
1705-9247
Interface PC Board (1)
4
1705-9031
Chassis (1)
5
1706-3447
Pump Assembly (1)
6
1705-9494
Nicad Battery Pack
or
1705-9312
Alkaline Battery Pack
7
1706-0716
Case Gasket (1)
8
1706-2498
Filter Housing (1)
9
1705-8157
Dust Filter/Water Stop (1.2 Micron) (1)
10
1705-7118
Speaker (1)
11
1705-8660
Speaker Gasket (1)
12
1705-8140
Speaker Water Barrier (1)
13
1706-0674
Sensor Manifold (1)
14
1702-8374
External Alarm Jack (1)
15
1705-0295
RFI Screen (1)
16
1706-1284
Case Top (1)
17
1705-8306
Key Pad (1)
18
1706-0500
Alarm Lens (1)
19
1705-8652
Alarm Lens Gasket (1)
The following items numbers refer to the exploded view
drawing on pages 24 and 25.
9.
R
EPLACEMENT
P
ARTS
WARNING: When removing a sensor from
service, the appropriate sensor plug
must be installed to ensure proper
instrument operation. Toxic/O
2
Sensor Plug P/N 1704-6947
Combustible Sensor Plug
P/N 1706-2647
8.4.1 TOXIC/OXYGEN SENSOR REMOVAL/
REPLACEMENT
• To remove a toxic or oxygen sensor, grasp the sensor
and lift it straight up. Use care to avoid bending the
sensor pins.
WARNING: When removing a toxic sensor for
storage, connect a shorting wire to
the two pins as shown.
• New toxic sensors are shipped with a shorting wire
attached. Remove the shorting wire from the new
sensor.
• Immediately install the sensor in the instrument. Never
apply pressure to the area inside of the black O-Ring seal
at the top of the sensor.
8.4.2 COMBUSTIBLE SENSOR REMOVAL/
REPLACEMENT
• To remove the combustible sensor, grasp the sensor and
lift it straight up.
• Press the new sensor firmly into the sockets on the
PC board.
8.5 C
HANGING THE
I
NTERNAL
F
ILTER
If the internal dust filter/water stop becomes blocked, a
PUMP FAULT alarm will result. The filter must be
replaced immediately before continuing operation. To
remove and replace the internal dust filter/water stop from
the instrument:
• Unscrew and remove the knurled sample inlet fitting from
the instrument as shown in Figure 6.
• Grasp the dust filter/water stop from the instrument and
pull straight out to remove.
• Replace the dust filter/water stop making sure that the
larger diameter opening of the filter is facing inward.
• Replace the knurled sample inlet fitting.
NOTE: The ATX612 will remain in a PUMP FAULT
condition if operation is attempted with the dust
filter/water stop removed.
!
!
Shorting Wire
Figure 6