Personal Single Gas Monitor 

 

Activation 

 Operation 

 Troubleshooting 

 
 

 

 

 

 

 
 
 

Part Number:

 17120932 

Revision:  

1.6 

Release Date:

  June 3, 2006 

For foreign language versions of 

this manual, go to www.indsci.com.

Содержание GASBADGE PLUS

Страница 1: ...Personal Single Gas Monitor Activation Operation Troubleshooting Part Number 17120932 Revision 1 6 Release Date June 3 2006 For foreign language versions of this manual go to www indsci com...

Страница 2: ...tion Section 15 Zeroing Failed 16 Zero Passed 17 Calibration 18 Days Since Calibration 20 Event Log 21 Initiate Print 21 Low Alarm Setpoint 22 High Alarm Setpoint 23 Cal Gas Setpoint 24 Set Display Ty...

Страница 3: ...M and for H2S performance over the range of 0 100 PPM and 2 EN50104 for O2 performance sensor over the range of 0 25 The GasBadge Plus is certified for use within ambient temperature range of 40o C to...

Страница 4: ...nitor 1 17120908 Belt Clip 1 17120932 Manual 1 17124033 Cal Cup 1 17093659 Urethane Tubing 1 17121385 Quick Start Guide After unpacking if any listed item is missing contact either your local distribu...

Страница 5: ...s Days and Hours Count down GasBadge Pro Only GasBadge Pro Only GasBadge Pro Only Alarm Auxiliary Display Confidence and Warning Icons Peak Value Zeroing Battery Status Apply Gas Calibration Security...

Страница 6: ...alarm indicator test the instrument will display the software version IMPORTANT Once the GasBadge Plus is activated it cannot be turned off NOTE Any button press activates the backlight for a period o...

Страница 7: ...Operating Quick Start Guides Basic Operating Flowchart Gas Monitoring Screen Battery Life Remaining Peak Reading To Security Clear Peak Loop To Print and Backlight Hold for 3 seconds 1 Minute Months D...

Страница 8: ...ssed Days Since Calibration Calibration Calibration Passed Calibration In Process Days Since Calibration IMPORTANT Calibrate the GasBadge Plus using a flow rate of 0 5 LPM IMPORTANT Before zeroing ens...

Страница 9: ...Confidence Indication Software Version Instrument Test Screen Toggle Value To Beginning of Configuration Mode Calibration Set Security Code D Total Alarm Time C Increment Value Edit Value Edit Value L...

Страница 10: ...eens will force a transition back to the Gas Monitoring screen The alarm screen is indicated on the display by the Alarm Indicator and either the Up Indicator or Down Indicator which designate a low o...

Страница 11: ...lay and the Checkmark Indicator if appropriate Battery Life Remaining Screen Months After 30 seconds this screen will timeout and return to the Gas Monitoring screen Pressing the Mode button at this s...

Страница 12: ...itiate or Calibration screen for O2 30 second timeout Go to Gas Monitoring screen Gas alarm event Go to Gas Monitoring screen With one minute of battery runtime remaining the Battery Failure screen ap...

Страница 13: ...toring screen Pressing the Enter button on the Peak Gas Reading screen will clear the peak value On O2 instruments the depletion peak will be cleared to 20 9 NOTE The GasBadge Plus retains the maximum...

Страница 14: ...calibration the instrument will return to the Gas Monitoring screen if no buttons are pressed for 30 seconds Security Code The Security Code screen is the first screen in the configuration mode The de...

Страница 15: ...by company policy or local regulatory agencies If an instrument fails a function or bump test or if it is dropped submerged or appears damaged a full calibration is recommended The safest approach is...

Страница 16: ...In Process Mode and Enter buttons held 3 sec Go to Gas Monitoring screen 30 second timeout Go to Gas Monitoring screen Zeroing Failed The Zeroing Failed screen is displayed as a zero icon and a flash...

Страница 17: ...r buttons held 3 sec No response Zero Passed The Zero Passed screen shows that the zero process completed successfully After 5 seconds the Calibration screen is displayed Zero Passed Screen Actions Av...

Страница 18: ...g the Mode button displays the Days Since Last Calibration screen If Cal Failed see Actions Available from Cal Failed Screen chart Calibration Screen Calibrating with Gas Cylinders Toxic Sensors or Ze...

Страница 19: ...ode and Enter buttons held 3 sec Go to Gas Monitoring screen 5 second timeout Go to Gas Monitoring screen Sample Cal Passed 9 and Cal Failed Screens Actions Available from the Cal Passed 9 Screen Cal...

Страница 20: ...he Days Since Calibration screen indicates the number of days that have passed since the last calibration The Days Indicator is active to designate this number as days and the Gas Bottle indicator is...

Страница 21: ...ilable from Industrial Scientific or it can be downloaded to a PC via a DataLink or CalPlus accessory product Initiate Print The Initiate Print screen allows you to print the last 15 alarm events to a...

Страница 22: ...resent threshold value All characters of the primary display will blink to indicate that they are ready to change The threshold is incremented by pressing the Enter button Simple momentary presses wil...

Страница 23: ...the value by 1 unit Holding the Enter button for an extended time increments the value much faster High Alarm Setpoint Screen Upon reaching the maximum value which is sensor dependent the display roll...

Страница 24: ...ndicate that they are ready to change The threshold is incremented by pressing the Enter button Simple momentary presses will increment the value by 1 unit Cal Gas Setpoint Holding the Enter button fo...

Страница 25: ...e sensor type on the auxiliary display along with the up and down arrows to indicate that the display type can be modified On the Sensor Type Only screen the sensor type will move to the primary displ...

Страница 26: ...at it is ready to change The Enter button will toggle the rightmost character between 0 and 1 The Mode button will cause a transition to the Set Security Code screen Set Confidence Indication Actions...

Страница 27: ...ode button displays the Total Alarm Time screen Set Security Code Screen Actions Available from the Set Security Code Screen Action Response Mode button press Accept current code then go to the Total...

Страница 28: ...oring screen Software Version On this screen the software version and revision are displayed The auxiliary display will indicate the software build number as an integer from 0 to 99 preceded by the le...

Страница 29: ...will time out back to the Instrument Test screen Mode and Enter buttons held Go to Gas Monitoring screen 30 second timeout Go to Gas Monitoring screen DataLink Optional Accessory The GasBadge DataLin...

Страница 30: ...ted using IrDA through the instrument s infrared communications port Using this communications port the calibration station performs bump tests and calibrations of the instrument The CalPlus saves the...

Страница 31: ...either high or low alarm respectively Sample High Alarm Screen Battery Failure Three levels of low battery warnings exist to let you know that the instrument is nearing the end of its life At seven da...

Страница 32: ...permanently in preparation for returning it to Industrial Scientific for repairs WARNING If this error occurs you must shut down the personal monitor and return it to Industrial Scientific for servic...

Страница 33: ...Nylon carrying case 18106419 2 unit nylon carrying case 17124033 Calibration cup Cal Cup 18106260 GasBadge DataLink 18106344 0X2 CalPlus Calibration Station 18106344 1X2 CalPlus Calibration Station w...

Страница 34: ...ay for direct gas readings Backlight for low light conditions User selectable option for direct gas reading mode or display just gas type Peak reading indication Alarms User selectable low and high al...

Страница 35: ...ppm 5 0 15 20 to 50 4 to 122 15 90 O2 0 30 vol 0 1 0 55 0 85 20 to 50 4 to 122 0 99 SO2 0 150 ppm 0 1 ppm 5 0 15 40 to 50 40 to 122 15 90 1 For any given sensor reading the accuracy shall be taken to...

Страница 36: ...or consequential damages including loss of profit or loss of use arising out of the sale manufacture or use of any products sold hereunder whether such claim is pleaded in contract or in tort includi...

Страница 37: ...15 17 Cal Passed 18 Cal Cup 4 18 33 calibrating with gas cylinders 18 calibration 3 14 18 20 24 30 35 cup 4 18 33 failed 18 flow rate 18 recommendations 15 gas 23 24 33 gas concentration setpoint 23 2...

Страница 38: ...port 10 21 print 10 21 L LEDs 6 limitation of liability 36 lithium battery 3 logger 34 Low Alarm Setpoint screen 21 low battery indication warning 11 31 M Mode button 6 10 17 19 29 monitoring screen 1...

Страница 39: ...on screen 6 sulfur dioxide 33 T temperature 3 35 range 34 Test Flash screen 29 test screen 29 Test screen 28 timeout 11 13 15 17 19 28 total alarm time 27 36 toxic gas sensor 36 toxic sensors 10 15 tr...

Страница 40: ...ial Scientific Australia Pty Ltd Level 5 113 Wicks Road North Ryde NSW 2113 Australia Phone 61 2 8870 3400 Fax 61 2 8870 3444 e mail australia indsci com ISCn Service Center T12B 4 5 Floor 615 Ningqia...

Страница 41: ......