Page 11
VaxArray® Imaging and Analysis System Operation Manual
CF-0159 R004
Troubleshooting, Cleaning, and
.
Hardware and Peripheral Connectors
Multiple connection options, including VGA and USB, are provided to connect the provided mouse,
keyboard and monitor along with any other hardware or peripherals.
Imaging Tray Open Button
In the event of a system failure or other instance in which the imaging tray cannot be opened via the
software interface, this button may be used as outlined below:
•
Shut down the system and switch the main power switch off.
•
Switch the main power on again.
•
Press the Imaging Tray Open button on the rear panel while pressing the power button on the
front panel.
•
The imaging door will open and the imaging tray will extend.
Imaging Tray and Slide Holder
The four slide holder should be placed into the imaging tray with the A1 engraved on the slide holder positioned
toward the top left near the A1 of the imaging tray. The slide(s) are to be placed in the 4 slide holder with the
first slide in the left-most
“Mount 1”
position with the A1 well of the slide located in the top left (
Note that if you are performing a standard VaxArray analysis that provides a quantitative (
μ
g/mL) result, the
leftmost slide in Mount 1 should contain the standards comprising the calibration curve in slide wells A1
through H1.
Figure 3: Imaging Tray & Slide Holder.
A1 Location