  B. Regarding the InBody Test

    Some of the more common clinical questions are answered below. If there are additional questions or more clarification is

    

desired, please contact InBody.

Question

Answer

• 

Must socks or stockings

  be removed for the 
  InBody Test?

• 

Bare skin contact is essential for analysis using the BIA method. Socks or 

  stockings may cause a varying degree of distortion in the results. Socks or 
  stockings must be removed to obtain accurate data.

• 

Is it okay to wear accessories

  (jewelry, watch, rings, etc) or 
  metal objects while taking 
  the InBody Test?

• 

The ideal condition for the analysis is to simply stand with no clothes and  

  no accessories. However, this may not always be possible. 
  Therefore, we recommend that the examinee remove as many clothing items
  and accessories as possible that may affect the weight.

• 

Who cannot take the

  InBody Test or will have
  difficulties taking the
  InBody Test?

• 

Individuals with implanted medical devices such as pacemakers, or essential

  support devices such as patient monitoring systems, must not use this

  equipment. The currents will flow through the body during the

  test, which may cause malfunctioning of the device or endanger lives.

• 

Children, amputees, or the elderly may have trouble testing if they cannot

  hold the hand electrodes or stand still on the foot electrodes.

• 

Can a person with metal 

  implants in their body take 
  the InBody Test?

• 

The ideal test circumstance is when the examinee does not wear anything 

  metallic. Individuals with metallic implants may have skewed test results 
  due to the conductivity of the metal affecting the results.

• 

As the weight of clothes and other items affects the results of the body 

  composition analysis, it is strongly recommended to take off any heavy 
  clothing or metallic items. Except for its minor effect on overall body weight, 
  jewelry does not affect the body composition analysis, as the contact point 
  of the InBody270 are hands and feet.

• 

I have limited mobility and

  cannot maintain the proper
  posture for the InBody Test.
  How can I still be tested?

• 

It is impossible to test if an individual cannot maintain contact with the

  hand or foot electrodes. InBody has a line of products that can perform body
  composition analysis on bedridden examinees, which allow these patients
  to stay in bed. For more information, please contact InBody.

• 

Is the electric current 

  harmful to the body?

• 

The bioelectrical impedance impedance method uses safe low level currents

  that is not harmful to the body. The safety of the InBody has been tested 
  and proven. The InBody products have been approved for medical use by  
  the CE and all over the world. Many medical institutions around the world 
  are actively using the InBody.

• 

How often should I take

  the InBody Test?

• 

Individuals who are undergoing any programs that may affect their body

  composition are strongly recommended to have the InBody Test done every two
  to four weeks.

• 

Consistent testing will allow individuals to track and monitor their progress 

  over time.

• 

What are the precautionary

  steps to ensure the accuracy
  of the InBody Test?

• 

Please refer to ‘A. Precautionary Steps’ in section ‘II. InBody Test’ in this 

  User’s Manual.

Содержание InBody270

Страница 1: ...User s Manual USA...

Страница 2: ...or damages incidental or consequential that occur by not complying with the content of the User s Manual InBody Co Ltd reserves the right to modify the appearance specifications etc of the InBody270 t...

Страница 3: ...B Test Instructions C Test Posture III Transportation and Storage A Cautions during Transportation B Repacking Instructions C Transportation and Storage Environment IV Frequently Asked Questions FAQ...

Страница 4: ...onents Please make sure all of the following components are present Please inspect each component of the InBody270 for defects prior to installation InBody270 Adapter DC 12V 3 4A 1 EA Power cord 1 EA...

Страница 5: ...e should be disposed of according to relevant laws and regulations DANGER NOTE UTION Caution Si vous avez des probl mes lors de l installation de votre InBody270 veuillez contacter InBody pour obtenir...

Страница 6: ...an uneven surface may cause the examinee to fall down Test results may also be inaccurate Never clean the hand and foot electrodes with liquid spray or detergent directly The equipment may corrode an...

Страница 7: ...er capacity Do not disassemble or modify the equipment including internal parts without written consent from the manufacturer This may cause electric shock or injury product malfunction inaccurate res...

Страница 8: ...nBody270 est soumis des interf rences lectriques N installez pas l InBody270 pr s d quipements qui g n rent des interf rences lectriques tels que les lampes fluorescentes les gros quipements moteur CA...

Страница 9: ...9 5 Flip the power switch to turn on the InBody270 InBody Power switch...

Страница 10: ...hing on top of the footplate Please do not stand on the footplate or place objects on the footplate 2 Press the Administrator Menu button on the screen which appears when no one is on the footplate 3...

Страница 11: ...st results as an Excel file on a USB Thumb Drive Exported test results can be viewed as an Excel file on a computer 08 Data Backup Restore Combine Back up InBody Test results to a USB Thumb Drive rest...

Страница 12: ...value on the InBody Example Workout clothes at the gym are approximately 0 2kg most examinees are assumed to be wearing workout clothes so the examiner may adjust the set value to 0 2kg 20 Reference S...

Страница 13: ...ng the InBody Test 02 Results Sheet does not print View the troubleshooting checklist when the Results Sheet does not print with the printer connected to the InBody 03 Weight is not being measured Vie...

Страница 14: ...the Administrator Menu from 10 Results Sheet Types through 15 Printing Alignment 2 Thermal Printer Connect thermal printer to the InBody270 to print Thermal Results Sheet For best results the InBody2...

Страница 15: ...l Connect If the stadiometer is connected to the InBody270 the stadiometer icon will appear on the top left corner of the screen when no one is on the footplate 4 Barcode Reader If a barcode reader is...

Страница 16: ...nel of the InBody270 Then connect the other end of the serial cable to the serial port of the SD400 3 Connect your devices such as the stadiometer and blood pressure monitor to the ports on the SD400...

Страница 17: ...ANGER NOTE ON Attention Veillez ne pas tordre les poign es des lectrodes main ou ne pas les faire pivoter dans le sens inverse au del de ses limites Ne prolongez pas les poign es des lectrodes main au...

Страница 18: ...tre supervis s ou assist s lorsqu ils essaient de faire le test sur l InBody Apr s qu un individu afflig d une maladie contagieuse ou d u e infection ait effectu le test sur l InBody utilisez un InBo...

Страница 19: ...de composition du corps mais le volume d urine et d excr ments est inclus dans la mesure du poids Cela affecte la pr cision des r sultats du test Ne faites pas de sport avant le test Des exercices in...

Страница 20: ...ofessional Mode Professional Mode An examiner is present and guiding the examinee through the InBody Test Self Mode The examinee takes the InBody Test following the instructions that are displayed on...

Страница 21: ...posture to take the test Refer to C Test Posture for the proper posture 5 The InBody Test begins 6 When the test is completed the results will be shown on the screen Professional Mode Self Mode Profe...

Страница 22: ...e with all of the electrodes The thighs should not touch Step on the footplate barefoot Hold the hand electrode so that the 4 fingers wrap the surface of the bottom hand electrode while placing the th...

Страница 23: ...lock button lower the upper part by folding it down 4 Put the InBody into the packing box Place the packing pad over the equipment and tape up the packing box DANGER NOTE CAUTION Caution Always use t...

Страница 24: ...roblem cannot be resolved by consulting the Troubleshooting section please refer to the possible solutions below Problem Solution My InBody270 does not turn on Insert the power plug completely into a...

Страница 25: ...tart the InBody The InBody stores the last touchscreen alignment and will automatically recall the previous touchscreen settings The InBody can also recognize if its touchscreen alignment is off scree...

Страница 26: ...viduals with metallic implants may have skewed test results due to the conductivity of the metal affecting the results As the weight of clothes and other items affects the results of the body composit...

Страница 27: ...he 4 fingers wrap the surface of the bottom hand electrode while the thumb is placed on the oval electrode Body Connects the upper part of the equipment to the lower part LCD screen Shows each stage o...

Страница 28: ...akes contact with this electrode by stepping with the heel of their foot Footplate This is connected to the scale which measures the examinee s weight Level Indicator Indicates the current horizontal...

Страница 29: ...Stadiometer and SD400 LAN port 10T Base Used for connecting the InBody270 to Lookin Body installed on a computer The InBody270 can be connected to Lookin Body installed on a computer using one of the...

Страница 30: ...contact a qualified technician or your local representative DANGER NOTE UTION Advertisement Les quipements externes destin s la connexion de l entr e du signal de la sortie du signal ou d autres conne...

Страница 31: ...collection point for the recycling of electrical and electronic equipment By ensuring this product is disposed of correctly you will help prevent potential negative consequences for the environment a...

Страница 32: ...DANGER Do not use this equipment with electrical medical devices such as a pacemaker Ne pas utiliser cet quipement avec des appareils m dicaux lectriques comme un stimulateur cardiaque CAUTION Do not...

Страница 33: ...Trunk Right Leg Left Leg Body Composition History Weight Skeletal Muscle Mass Percent Body Fat Body Fat Lean Body Mass Control Lean Body Mass Basal Metabolic Rate Results Interpretation QR Code Imped...

Страница 34: ...LCD Internal Interface Touchscreen Keypad External Interface RS 232C 1EA USB HOST 2EA USB SLAVE 1EA LAN 10T 1EA Bluetooth 1EA Wi Fi 1EA Compatible Printer Laser Inkjet Printers Printers recommended b...

Страница 35: ...contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transie...

Страница 36: ...rom fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked wi...

Страница 37: ...37...

Страница 38: ......

Страница 39: ......

Страница 40: ...www inbodyusa com 2015 InBody Co Ltd All rights reserved BM USA F9 C 151111...

Отзывы: