Maintenance

Use the arrow keys to select 

Add Sensor

.  When 

complete, the RDO-X Stainless Steel will appear 

on the list.

The display will now show readings from the 

RDO-X Stainless Steel.

Add Sensor.

Read Display.

How to Install and Replace the RDO Cap

Unscrew and remove 

the guard.

Slide the RDO 

cap away from the 

instrument. Do not 

twist cap.

Align the flat edge 

inside the cap with 

the flat edge on the 

instrument. Slide the 

new cap into place.

Replace the guard 

and calibrate the 

instrument.

Align 

with cap