16
4-6. Sensitivity Setting of AC and DC Output
Fig 13 will appear as this icon is selected.
Japanese
English
Fig 13 AC and DC Output Sensitivity Setting Page
This function set the full-scale value to the AC Output Voltage(1V). Display in the Figure 14 will appear
once you press “Set.”
You can set the level of velocity, acceleration, and displacement per 1V. In the Figure 14, each level is set
as follows:
Velocity: 1V is equivalent to 1500mm/s.
Acceleration: 1V is equivalent to 450m/s
2
.
Displacement: 1V is equivalent to 75μm.
The physical amount of output signal of AC OUT is equivalent to the physical amount designated in
the mode setting. You can select the value by pressing the button. The value will be switched as follows:
Velocity: 1500
→
150
→
15
→
15000
Acceleration: 45
→
4.5
→
450
Displacement: 75
→
7.5
→
0.75
→
0.075