53
Troubleshooting Wireless Connection between
OPTIMIZER Smart Mini IPG and Vesta Charger
If you experience issues with establishing a wireless
connection between the OPTIMIZER Smart Mini IPG and
the Vesta Charger, try the following:
Whenever the Vesta Charger is not being used to
charge the OPTIMIZER Smart Mini IPG, place it
in an area that is frequented by the patient (e.g.,
bedside table in the bedroom), connected to its
AC Adapter, and the AC Adapter plugged into the
wall outlet. This will ensure regular
communications between the OPTIMIZER Smart
Mini IPG and the Vesta Charger.
Remain stationary during the charging or data
transfer process.
Decrease the distance between the devices.
Move the devices so they share line of sight.
Move the devices away from other devices that
may be causing interference.
Do not operate other wireless devices (i.e.,
programmers for other devices, laptop, tablet,
mobile phone, or cordless phone) at the same
time.
Wait a few minutes and try connecting again.
NOTE:
Wireless communications equipment, such as
wireless home network devices, mobile and cordless
telephones, and tablets, could affect the quality of the
wireless connection.