34
Explorer
Pro
3.16
Printing Data (Cont.)
Span Calibration Printout
When performing a Span calibration, a printout is automati-
cally made after the calibration mass is placed on the pan and
the
button is pressed.
Linearity Calibration Printout
When performing a Linearity calibration with GLP turned on, a
printout is automatically made after the calibration mass is
placed on the pan and the
button is pressed.
Calibration Test Printout
When performing a Calibration Test with GLP turned on, a
printout is available. When the display indicates the mass value
to be placed on the pan, the balance the automatically displays
the calibration weight required and the
button is pressed.
- - - - - CAL TEST - - - - - -
7/01/97
1:00:00 PM
Bal Id 1234
Cal:
4000.0g
Act:
4000.4g
Dif:
0.4g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - SPAN CAL - - - - - -
7/01/97
1:00:00 PM
Bal Id 1234
Cal:
4000.0g
Old:
4000.0g
Dif:
0.0g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - LIN CAL - - - - - -
7/01/97
1:00:00 PM
Bal Id 1234
Cal:
4000.0g
Old:
3999.4g
Dif:
0.6g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
SAMPLE PRINTOUTS