The parameters are activated dynamically with setting the "Average parameter".
Use the / buttons to navigate between
the individual parameters. The lower right
area shows the current value of the selected
parameter. To change the value, press the
key .
The value now appears in the display. Use
the / buttons to adjust the value and
press the button to accept. You can also
leave the entry without changing the value.
Repeat the procedure for all parameters to be
adjusted.
If you have adjusted all parameters as
desired, select "SAVE" and confirm with the
key . Now the parameters are written back
into the probe and are now active.
You can leave the menu item "Average
Parameters" with the button
without
saving. Than all changes will be lost!
4.5.6 Basic Balancing
At the exchange of a sensor head, due to deviating cable lengths, it may be necessary to
perform a basic balancing in air. Hereby, the moisture measurement value of the probe is re-
aligned to the correct „Zero Value“.
Actuate the button
in order to start the basic
balancing. The balancing will be performed
subsequently.