The LEDs will also degrade in intensity during their lifetime. To ensure that all measurements including the auto-

generated and standard illuminants, are correct, you must perform a spectral calibration regularly.  

You must also consider the degradation of the LED when saving self-defined presets. If you save a preset with 

LED channels that use its maximum intensity, the possibility exists that this intensity cannot be reached after the 

burn-in time or the long-time degradation of the LED. In this case, you will get a warning message from the iQ-

LED control software. 

During the first 500-600 working hours, we recommend performing a spectral calibration every 50 operating 

hours. 

After the first 500-600 operating hours, a calibration of every 150 working hours suffices. 

Other factors that indicate the need for a spectral calibration include unsatisfactory illuminant generation, the 

aberration of the intensity values, or a spectral curve that does not fit with the predefined standard illuminants of 

the corresponding preset.  

 

The spectrometer works correctly 

 

The spectrometer settings are correct 

 

All LED channels work correctly 

 

The dark measurement is correct 

 

Your measurement environment is correct 

 

Your ambient temperature is correct 

How to perform the spectral calibration is described in the iQ-LED control software manual. 

4.4  Low-Intensity use 

When using your system with very low intensity, the spectral measurement values will start to fluctuate. The 

lower the intensity, the higher the fluctuation. The generated light is still stable up to a certain point. The 

fluctuation of the values is caused by the noise of the spectral measurement of the internal spectrometer. The 

higher the noise influence, the lower the light intensity. An approximate value will no longer be possible when 

using standard illuminants with an intensity lower than 25 lux.  

4.4.1  Low light in combination with a neutral density chart 

You can use a neutral density filter chart and activate the neutral density correction in the iQ-LED control software. 

The light source will run at a higher intensity while still having low intensity behind the chart plane. Please use the 

'Density Filter Compensation' in your iQ-LED software to get the corresponding intensity behind the filter. Please 

see chapter 3.1.3 in the software manual for details. 

 

Содержание LE7 E

Страница 1: ...ineering GmbH Co KG Im Gleisdreieck 5 50169 Kerpen Germany T 49 2273 99 99 1 0 F 49 2273 99 99 1 10 www image engineering com LE7 x2 x4 x6 E IR User manual for the LE7 product line x2 x4 x6 E IR January 14 2020 ...

Страница 2: ...Overview display and ports 5 3 2 Connecting the hardware 7 4 OPERATING INSTRUCTIONS SOFTWARE 8 4 1 Requirements 8 4 2 Software installation 8 4 3 Starting the system 8 4 3 1 Spectrometer settings 8 4 3 2 Spectrometer calibration 8 4 3 3 iQ LED valibration 8 4 4 Low intensity use 8 4 4 1 Low light in combination with a neutral density chart 9 4 4 2 Using single iQ LED modules for low light 10 4 4 3...

Страница 3: ...pectrometer is also included with the device Only suitable for indoor use Place your system in a dry constantly tempered environment without light interference The optimal ambient temperature range is 22 to 26 degrees Celsius The maximum ambient temperature range is 18 to 28 degrees Celsius The optimal system temperature range displayed in the software user interface is between 35 and 50 degrees C...

Страница 4: ...ook directly into the sphere or light source when using high intensities or sequences with a low response time Do not open the device without instructions from the Image Engineering support team or when connected to the power supply 2 GETTING STARTED 2 1 Scope of delivery o Integrating sphere with a dual slot for D280 size charts o Spectrometer built in o Power cord o USB cable o Control software ...

Страница 5: ... the first row of both iQ LED devices in the top and bottom borders of the image 5 From this position increase the distance of your camera to the LE7 so that both LED rows will move out of the image the minimum increase of distance should be 45 mm 6 Insert your test chart again and proceed with your test When using a diffuse test chart for shading or uniformity measurements you can move your camer...

Страница 6: ... saved light sequence with different illuminants it is possible to save one sequence on the device Use the power button to turn on and off the light There are three pre stored illuminants on your device the intensity of each illuminant is shown in the acceptance protocol of your device 1 illuminant A default illuminant 2 illuminant D50 3 illuminant D75 Note To store your generated illuminants or s...

Страница 7: ...ur test setup For example with an iQ Trigger see 2 1 optional equipment 3 2 Connecting the hardware Connect the power cord to the power outlet on your LE7 Connect the LE7 USB cable to your PC and turn on the LE7 the power switch is next to the power outlet Then connect the USB cable from the spectrometer calibration device to your LE7 The system will install the spectrometer and iQ LED driver on y...

Страница 8: ...oftware see iQ LED software manual automatically generates the optimal spectrometer settings for your lighting conditions after pressing the auto detect button For special applications it is also possible to set the spectrometer settings manually If you have further questions please contact the Image Engineering support team 4 3 2 Spectrometer calibration Your spectrometer comes fully NIST traceab...

Страница 9: ...r works correctly The spectrometer settings are correct All LED channels work correctly The dark measurement is correct Your measurement environment is correct Your ambient temperature is correct How to perform the spectral calibration is described in the iQ LED control software manual 4 4 Low Intensity use When using your system with very low intensity the spectral measurement values will start t...

Страница 10: ...tion chart should correspond to an intensity of approximately 0 25 lx when using only the iQ LED module with the ND filter To get a rough idea of which intensities will appear when using only the ND module you can use a Density Filter Compensation value of 2 68 in the iQ LED SW Please see chapter 3 1 3 in the software manual for details Please note that it is impossible to measure such low intensi...

Страница 11: ...mage Engineering Seite 11 von 11 6 TECHNICAL DATASHEET See annex for the technical data sheet It can also be downloaded from the website of Image Engineering https image engineering de support downloads ...

Отзывы: